OpenTrons-OT2 single-channel pipette specifications [33].

Pipette Volume Accuracy %D Accuracy μL Precision %CV Precision μL
P20 Gen2 1 ±15% 0.15 μL ± 5% 0.05 μL
P20 Gen2 10 ±2% 0.2 μL ± 1% 0.1 μL
P20 Gen2 20 ±1.5% 0.3 μL ± 0.8% 0.16 μL
P300 Gen2 20 ±4% 0.8 μL ± 2.5% 0.5 μL
P300 Gen2 150 ±1% 1.5 μL ± 0.4% 0.6 μL
P300 Gen2 300 ±0.6% 1.8 μL ± 0.3% 0.9 μL