Table 2.

Hounsfield unit (HU) values of the Contrast phantom measured from CBCT M, N1 ms-CBCT, ms-CBCT, and clinical MDCT.

CBCT M N1 ms-CBCT ms-CBCT MDCT Nominal value
HU HU HU HU HU
Acrylic 160.2±42.3 66.7±44.0 87.5±41.2 123.8±27.1 120
Air 867.6±44.1 849.7±30.2 973.6±19.2 1023.3±0.9 1000
Macor 1744.2±70.5 1583.3±82.9 2105.2±79.8 2423.3±42.1
LDPE 25.8±44.5 103.3±41.5 126.3±34.1 119.6±22.0 95
Water 87.7±44.0 27.6±55.5 11.5±47.7 15.4±24.7 0
RMSE 316 385 145

Note: (1) Macor is a machinable glass ceramic composed of ~55% fluorophlogopite mica and 45% borosilicate glass. (2) The nominal values are provided by the standard CT ACR 464 phantom which is suggested for 120–130 kVp39,43 rather than the 90 kVp used in this study.

CBCT cone beam computed tomography, ms-CBCT multisource cone beam computed tomography, MDCT multidetector computed tomography, LDPE low-density polyethylene, RMSE root-mean-square error.