Table 2.

Radiation dose metrics and technical parameters overall and stratified by patient size category

Variable (units) Patient size category
All patients Small Medium Large
n = 748,846 n = 165,816 n = 360,850 n = 222,180
Mean Std Dev Mean Std Dev Mean Std Dev Mean Std Dev
Patient diameter (mm) 309 (45) 252 (17) 301 (16) 363 (28)
CTDIvol (mGy) 13 (7.2) 8 (3.3) 12 (4.9) 20 (7.8)
DLP (mGy-cm) 798 (546) 447 (283) 701 (404) 1217 (629)
Effective dose (mSv) 13 (8) 9 (6) 12 (7) 16 (9)
mAs 181 (89) 120 (55) 166 (68) 251 (94)
Pitch* 1.0 (0.3) 1.0 (0.2) 1.0 (0.3) 1.0 (0.3)
Effective mAs 183 (95) 118 (54) 166 (67) 260 (106)
kV (kilovoltage) 119 (7.5) 116 (8.0) 118 (7.4) 121 (6.8)
Scan length (cm) 48 (9.2) 45 (8.8) 48 (9.0) 51 (9.2)
Slice thickness (mm) 3.7 (1.6) 3.7 (1.7) 3.7 (1.6) 3.7 (1.6)
Number of phases 1.3 (0.7) 1.3 (0.7) 1.3 (0.7) 1.3 (0.6)
Calculated size-adjusted variables
Size-adjusted CTDIvol (mGy) 12 (4.6) 11 (5.0) 12 (4.6) 12 (4.2)
Size-adjusted DLP (mGy-cm) 702 (389) 690 (448) 701 (388) 712 (338)

Abbreviations: CTDIvol volumetric computed tomography dose index, DLP dose length product, mAs milliampere-seconds, mGy milliGray, mSv millisieverts

*Table distance traveled in one 360° gantry rotation divided by beam collimation