Table 2.
Cumulative tube current values for the chest scan and information whether the scan was done with or without out-of-plane shield in localizer images (but outside the scanned volume). The difference between these two values, their relative difference, and location of the > 20% tube current increase along the z-axis (starting point of the scan is valued as 0 cm and the end of the scan 32 cm)
| Pitch | Cumulative tube current, no shield (mA) | Cumulative tube current, with shield (mA) | Difference (shield-no shield) (mA) | Relative difference (%) | z-location of the > 20% tube current difference (cm) | |
|---|---|---|---|---|---|---|
| GE Optima | 0.516 | 5065 | 8808 | 3743 | 74 | 29.1 |
| 0.984 | 11,302 | 24,004 | 12,702 | 112 | 27.3 | |
| 1.375 | 15,393 | 26,279 | 10,886 | 71 | 27.9 | |
| 1.531 | 15,818 | 50,218 | 34,400 | 217 | 23.1 | |
| GE Revolution | 0.508 | 14,336 | 19,662 | 5326 | 37.2 | 26.9 |
| 0.992 | 37,290 | 53,168 | 15,878 | 42.6 | 26.3 | |
| 1.375 | 53,312 | 71,015 | 17,703 | 33.2 | 25.5 | |
| 1.531 | 55,147 | 78,889 | 23,742 | 43.1 | 24.2 | |
| Canon One | 0.637 | 63,954 | 66,499 | 2545 | 3.98 | 29.9 |
| 0.813 | 69,547 | 76,739 | 7192 | 10.3 | 29.0 | |
| 1.388 | 121,837 | 143,648 | 21,811 | 17.9 | 27.1 | |
| Canon Prime | 0.637 | 56,247 | 66,242 | 9995 | 17.8 | 29.5 |
| 0.813 | 63,607 | 76,941 | 13,334 | 21.0 | 28.6 | |
| 1.388 | 109,280 | 167,543 | 58,263 | 53.3 | 22.3 | |
| Canon Prime SP | 0.637 | 181,542 | 187,897 | 6355 | 3.50 | 30.6 |
| 0.813 | 206,421 | 214,552 | 8131 | 3.94 | 30.9 | |
| 1.388 | 321,365 | 322,690 | 1325 | 0.41 | - | |
| Siemens Go.Up | 0.35 | 14,934 | 15,065 | 131 | 0.88 | - |
| 0.65 | 24,020 | 24,063 | 43 | 0.18 | - | |
| 0.95 | 36,181 | 36,230 | 49 | 0.14 | - | |
| 1.2 | 45,283 | 45,275 | − 8 | − 0.02 | - | |
| 1.5 | 57,605 | 57,402 | − 203 | − 0.35 | 31.7 |