Table 2.

Cumulative tube current values for the chest scan and information whether the scan was done with or without out-of-plane shield in localizer images (but outside the scanned volume). The difference between these two values, their relative difference, and location of the > 20% tube current increase along the z-axis (starting point of the scan is valued as 0 cm and the end of the scan 32 cm)

Pitch Cumulative tube current, no shield (mA) Cumulative tube current, with shield (mA) Difference (shield-no shield) (mA) Relative difference (%) z-location of the > 20% tube current difference (cm)
GE Optima 0.516 5065 8808 3743 74 29.1
0.984 11,302 24,004 12,702 112 27.3
1.375 15,393 26,279 10,886 71 27.9
1.531 15,818 50,218 34,400 217 23.1
GE Revolution 0.508 14,336 19,662 5326 37.2 26.9
0.992 37,290 53,168 15,878 42.6 26.3
1.375 53,312 71,015 17,703 33.2 25.5
1.531 55,147 78,889 23,742 43.1 24.2
Canon One 0.637 63,954 66,499 2545 3.98 29.9
0.813 69,547 76,739 7192 10.3 29.0
1.388 121,837 143,648 21,811 17.9 27.1
Canon Prime 0.637 56,247 66,242 9995 17.8 29.5
0.813 63,607 76,941 13,334 21.0 28.6
1.388 109,280 167,543 58,263 53.3 22.3
Canon Prime SP 0.637 181,542 187,897 6355 3.50 30.6
0.813 206,421 214,552 8131 3.94 30.9
1.388 321,365 322,690 1325 0.41 -
Siemens Go.Up 0.35 14,934 15,065 131 0.88 -
0.65 24,020 24,063 43 0.18 -
0.95 36,181 36,230 49 0.14 -
1.2 45,283 45,275  − 8  − 0.02 -
1.5 57,605 57,402  − 203  − 0.35 31.7