Published in final edited form as: Appl Opt. 2020 May 1;59(13):3985–3991. doi: 10.1364/AO.382408

Table 2.

Sources of Uncertainty and Their Impacts on the Index Uncertainty

Sources x Magnitude of Uncertainty u(x) Index Uncertainty ux(n)(105)

Peak position (0.7 to 5) arcseconds (1/10th of diffracted peak width FWHM) 0.6–4
Prism apex angle 0.15 arcseconds [including λ/120 (at 633 nm) RMS surface flatness error] 0.9
Index inhomogeneity 0.91 × 10−5 RMS for 10 mm thickness at λ= 3.39 μm 0.91
Encoder angle 0.15 arcseconds 0.17
Prism temperature 0.005°C 0.26–0.20
Wavelength calibration 0.05 nm 0.6–0.002
Alignment 0.05 arcseconds 0.08
N2 gas and air dispersion formulas 2 × 10−7 0.08
N2 gas temperature 0.02°C 0.01
N2 gas pressure 0.02 kPa 0.02