Skip to main content
. 2024 Jan 29;54(3):140–153. doi: 10.4070/kcj.2023.0252

Table 2. Comparison of electrocardiographic parameters between novel telemonitoring system using a single-lead ECG patch and the conventional telemonitoring system in inpatient setting.

Variables Novel telemonitoring system* (n=80) Conventional telemonitoring system (n=80) 95% limits of agreement ICC (95% CI) p value for reliability p value for difference
Monitoring time (minute) 1,408 (1,215–1,827) 1,408 (1,215–1,827) - - - -
Total QRS complex (No.) 69,635 (54,756–90,223) 66,601 (53,749–88,667) −14,573, 10,851 0.999 (0.998–0.999) <0.001 <0.001
Total atrial premature complex (No.) 99 (19–375) 100 (19–377) −181, 137 0.998 (0.997–0.999) <0.001 0.001
Total ventricular premature complex (No.) 5 (0–97) 5 (0–88) −308, 335 0.999 (0.998–0.999) <0.001 0.003
AF burden (%) 26.1 (10.1–100.0) 27.2 (10.0–100.0) −3.8, 4.3 0.995 (0.978–0.999) <0.001 0.116
Maximum HR (/min) 114.1±27.6 104.8±21.6 −37.8, 19.3 0.906 (0.853–0.940) <0.001 <0.001
Minimum HR (/min) 54.7±10.9 55.5±10.1 −6.6, 8.2 0.967 (0.949–0.979) <0.001 0.049
Average HR (/min) 72.2±12.1 71.7±12.0 −6.1, 5.1 0.986 (0.978–0.991) <0.001 0.151
Pause (second) 1.3 (1.2–1.7) 1.4 (1.3–1.8) −0.09, 0.17 0.997 (0.996–0.998) <0.001 <0.001

Data are mean ± standard deviation or median (interquartile range).

AF = atrial fibrillation; ECG = electrocardiogram; HR = heart rate; ICC = intra-class correlation coefficient.

*ThynC™, Seers Technology, Seongnam, Korea.

IntelliVue MX40 Patient Wearable Monitor, Philips, New York, NY, USA.

Among a total of 80 participants, AF was detected in 9 participants. Both the novel telemonitoring system using a single-lead ECG patch and the conventional system detected AF in the same 9 participants. The AF burden data was calculated from the 9 participants only.