Skip to main content
. 2024 Jan 29;54(3):140–153. doi: 10.4070/kcj.2023.0252

Table 3. Comparison of signal loss and noise between novel telemonitoring system using a single-lead ECG patch and the conventional telemonitoring system.

Variables Novel telemonitoring system* Conventional telemonitoring system p value for difference
Proportion of noise data (%) 0.3 (0.1–1.6) 2.4 (1.4–3.7) <0.001
Episodes of signal loss§ (N) 22 (2–53) 64 (22–112) 0.002
Minimum duration of signal loss (sec) 0.5 (0.5–0.6) 0.5 (0.5–0.6) 0.325
Maximum duration of signal loss (sec) 4.5 (2.5–5.1) 4.5 (4.1–4.9) 0.064

Data are presented as median (interquartile range).

*ThynC™, Seers Technology, Seongnam, Korea.

IntelliVue MX40 Patient Wearable Monitor, Philips, New York, NY, USA.

Most noise data were due to signal noise from motion artifacts or poor contact between electrodes and skin. The noise data excluded the episodes with signal loss.

§Most episodes of signal loss were due to transient loss of wireless connection due to participants wandering outside the effective range of the AP.

AP = access point; ECG = electrocardiogram.