. 2024 Mar 7;14(3):2640–2654. doi: 10.21037/qims-23-1319

Table 2. Patients by the automatic CMB detection software versus human rater.

Variables Automatic CMB detection software Radiologist P value
Total, n 265 265
True positive, n 145 139
False positive, n 27 5
False negative, n 3 9
True negative, n 90 112
Sensitivity (95% CI) (%) 98.0 (95.7–100.0) 93.9 (90.0–97.8) 0.656
Specificity (95% CI) (%) 76.9 (69.2–84.7) 95.7 (92.0–99.4) 0.212
Rule out patients without CMBs, n 93 117 0.016

P values are for the chi-squared test. CMB, cerebral microbleed; CI, confidence interval.