Table 2. Patients by the automatic CMB detection software versus human rater.
| Variables | Automatic CMB detection software | Radiologist | P value |
|---|---|---|---|
| Total, n | 265 | 265 | – |
| True positive, n | 145 | 139 | – |
| False positive, n | 27 | 5 | – |
| False negative, n | 3 | 9 | – |
| True negative, n | 90 | 112 | – |
| Sensitivity (95% CI) (%) | 98.0 (95.7–100.0) | 93.9 (90.0–97.8) | 0.656 |
| Specificity (95% CI) (%) | 76.9 (69.2–84.7) | 95.7 (92.0–99.4) | 0.212 |
| Rule out patients without CMBs, n | 93 | 117 | 0.016 |
P values are for the chi-squared test. CMB, cerebral microbleed; CI, confidence interval.