Table 1. Flow diverters available in Japan.

FDs Approved year Indications Structure Material Surface material
Pipeline Flex with Shield technology* 2015
  • ICA (petrous to supraclinoid), VA

  • Aneurysm size ≥5 mm

  • Length of neck ≥4 mm or D/N ratio <2

Mono layer (48 wires) Cobalt/chromium (36 wires)
Platinum/tungsten (12 wires)
MPC polymer
FRED** 2019 ICA, proximal ACA, proximal MCA, VA, BA
  • Aneurysm size ≥5 mm

  • Length of neck ≥4 mm or D/N ratio <2

Dual layer (66 wires) Nitinol (64 wires)
Tantalum (2 wires)
Bare metal
Surpass Streamline*** 2021 ICA (petrous to supraclinoid)
  • Aneurysm size ≥10 mm

  • Length of neck ≥4 mm or D/N ratio <2

Mono layer (72 or 96 wires) Cobalt/chromium Platinum/tungsten Bare metal

*Medtronic, **Terumo MicroVention, and *** Stryker.