Table 2. Subject impedance measurements at low currents with ECT electrodes.

Table includes Thymatron (pre, post and average measurements) measurements, impedance analyzer measurements at 1 kHz, and the lumped circuit values approximating impedance spectrum (>100 Hz data only) with the 1 kHz impedance of that circuit. EIS curve fit element parameters are for Trial 1. Note, impedance analyzer Trial 1 is for 1 V (10–1000Hz), Trial 2 is 1 V (1–1000Hz), and 100 mV (10–1000 Hz). IMP: Impedance.

Subject
ID
Thymatron IMP (Ω) IMP Analyzer 1 kHz (Ω) Circuit Values IMP
1 kHz(Ω)
Pre Post Avg. Trial 1 Trial 2 Trial 3 Rs (Ω) Rp (kΩ) Cp (nF)
201 1210 1290 1250 1253 1255 1242 522 9.89 175 1086
202 1690 1640 1665 1621 1630 1690 681 13.17 138 1386
203 1000 1140 1070 1102 1111 1089 506 5.52 210 960
204 1460 1520 1490 1485 1490 1467 508 13.44 138 1297
205 1670 1710 1690 1673 1716 1690 555 22.59 119 1475
206 1090 1090 1090 1033 1055 1040 449 4.24 219 906
207 850 770 810 808 808 804 372 2.84 294 810
208 1370 1370 1389 1389 1357 1376 557 6.70 160 1203
209 1020 900 939 939 904 980 432 2.57 263 806
210 940 830 885 838 818 862 374 3.97 277 723