TABLE 3.
Comparison of analytical enhancement factors (AEF) calculated for the surface‐enhanced Raman scattering (SERS) substrates in comparison to bare Al foil for the measurement of rhodamine 6G at 1064 nm excitation.
| Substrate | Wavelength (nm) | c R6G (ppm) | Peak area (a.u.) | AEF (103) |
|---|---|---|---|---|
| Al foil | 1064 | 1000 | 0.087 | – |
| AgNPL | 1064 | 0.1 | 0.096 | 11.1 |
| AuNP | 1064 | 1 | 0.039 | 0.4 |
| 4n SERS | 1064 | 0.1 | 0.040 | 4.6 |