TABLE 1.

Comparison of typical performance characteristics of commonly used mass spectrometers. Table generated using data from 1 , 11 , 12 , 13

Mass analyzer type a Resolving power (FWHM) [x103] Mass accuracy (ppm) m/z range (upper limit) [x103] Price
Quadrupole (Q) < 5 > 100 2‐4 Lower
Ion trap (IT) < 5 < 30 4‐20 Lower
TOF b 10‐60 0.5‐5 100 Moderate
Orbitrap 120‐1000 0.5‐5 20 Higher
FT‐ICR 100‐10 000 0.05‐1 30 High
a

TOF, Orbitrap and ICR include common hybrid configurations with Q or LIT as the first mass analyzer.

b

Values quoted for reflectron‐TOF instruments. Linear TOFs capable of greater m/z ranges (up to 500 000 in practice, and theoretically limitless) with lower resolving powers.