Table 3.

Comparison of IVUS vs CCTA:Separately 64-slice vs 256-slice (continuous variables)

Metric IVUS (n=686) CCTA (n=686) Correlation Coefficient (P-value*) P value of the correlation coefficient 64-slice vs. 256 slices**
Vessel area (mm2)
256-slice CCTA (n=334) 13.1±4.2 11.1±4.5 0.60 (p<0.0001) P=0.40
64-slice CCTA (n=352) 12.4±4.5 10.3±4.5 0.64 (p<0.0001)
Lumen area (mm2)
256-slice CCTA (n=334) 6.8±2.7 5.6±2.7 0.41 (p<0.0001) P=0.52
64-slice CCTA (n=352) 6.7±2.8 5.0±2.4 0.45 (p=0.002)
Plaque area (mm2)
256-slice CCTA (n=334) 6.1±2.9 4.9± 3.1 0.67 (p<0.0001) P=0.81
64-slice CCTA (n=352) 5.8±2.9 5.3±3.3 0.68 (p<0.0001)
MLA per artery (mm2)
256-slice CCTA (n=334) 5.0±2.0 4.1±1.3 0.61(p<0.0001) P=0.0004
64-slice CCTA (n=352) 4.0±1.2 3.0±1.4 0.41(p=0.24)
Average ESS (Pa)
256-slice CCTA (n=334) 2.6±1.4 2.9 ±2.1 0.41 (p<0.0001) P=0.23
64-slice CCTA (n=352) 2.7±1.7 3.3±3.7 0.33 (p=0.017)
Plaque 90° arc MIN ESS (Pa)
256- slice CCTA (n=334) 1.9± 1.4 2.1±1.8 0.32 (p<0.0001) P=0.48
64-slice CCTA (n=352) 2.0±1.5 2.8±3.2 0.27 (p<0.008)
Plaque 90° arc MAX ESS (Pa)
256-slice CCTA (n=334 3.3 ±1.6 3.8±2.5 0.49 (p<0.0001) P=0.19
64-slice CCTA (n=352) 3.4 ±1.7 4.6±4.4 0.41 (p=0.007)

Values are presented as mean± SD for continuous variables, n= number of analyzed segments, CCTA= coronary computed tomography angiography, IVUS=intravascular ultrasound, Min ESS= minimal endothelial shear stress, Max ESS= maximal endothelial shear stress, Average ESS=average endothelial shear stress

*

Correlation coefficient is different from r=0.25

**

Comparison of 2 correlation coefficients