Table S2:

Jumbo cylinder (JC) refilling capacity and refilling booster capacity of various PSA plants

PSA Plant Capacity (LPM) Refilling Capacity (JC/day)* Refilling Booster Capacity (kWh)
100 15 2.5
200 31 5
400 62 10
500 77 12.5
800 123 20
1000 154 25
1200 185 30
1500 231 35
2000 309 50
2500 386 60
3200 494 80

*The cylinder refilling capacity (jumbo cylinders (JC)/day) may vary marginally among various PSA plant models. In this study, the refilling capacity has been assumed to be 75% of the total PSA plant capacity. The refilling booster capacity (kWh) may vary marginally among various PSA plant models. JC – Jumbo Cylinders; PSA – Pressure Swing Adsorption; LPM – Litres Per Minute