Skip to main content
. 2022 Nov 15;96(1141):20220387. doi: 10.1259/bjr.20220387

Table 4.

Comparison of spatial resolution measurements

PET System (crystal size) 1 cm 10 cm 20 cm
FWHM FWHM FWHM
Axial Tangential Radial Axial Tangential Radial Axial Tangential Radial
RefleXion (4.1 × 4.1x 20 mm3) 4.2 4.3 4.3 5.2 4.8 5.2 7.0 5.8 7.2
Vereos (4 × 4 x 19 mm3) 4.2 N/A N/A 4.4 4.0 4.0 4.6 4.9 5.8
Discovery MI (3.95 × 5.3×25 mm3) 4.4 4.0 4.0 5.5 4.2 5.3 5.7 4.7 7.5
Vision (3.2 × 3.2 x 25 mm3) 3.5 3.5 3.6 4.3 3.9 4.5 4.4 3.5 5.8
Celesteion (4 × 4 x 4 mm3) 4.4 4.7 4.5 4.6 4.8 4.6 4.7 5.3 5.8