Skip to main content
. 2022 Nov 15;96(1141):20220387. doi: 10.1259/bjr.20220387

Table 6.

Image contrast and background variability for different sphere diameters (whole-body PET), with sphere-to-background ratio of 4:1

PET system Image Contrast % (Background Variability%)
10 mm 13 mm 17 mm 22 mm 28 mm 37 mm
RefleXion N/A 29.6 (14.8) 64.9 (12.4) 66.5 (10.3) 81.8 (8.8) 81.2 (8.3)
Philips-Vereos 38.4 (8.4) 61.3 (7.1) 65.9 (5.9) 68.5 (4.6) 83.4 (3.5) 86.4 (2.8)
GE-Discovery MI 51.7 (10.0) 61.5 (7.8) 66.2 (6.0) 81.3 (3.8) 86.6 (3.8) 90.0 (3.0)
Siemens-Biograph Vision 93.1 (6.4) 73.5 (5.0) 79.0 (4.0) 87.0 (3.1) 86.3 (2.7) 89.4 (2.2)
Toshiba- Celesteion 27.9 (6.2) 48.6 (5.3) 52.0 (4.6) 60.5 (4.1) 72.9 (3.9) 77.1 (3.8)