Table 1.

Calibration of graphite-core SSC in photon beams

Nominal Energy (TPR20,10) Depth cm Field size cm×cm SSD cm Dose rate Gy/min Actual dose cGy SSC ΔT K Calibration coefficienta
Gy/K Relative to Co-60
Co-60 (0.568) 5 10 × 10 60 1.4 141.0 0.0016728 842.6 1.000
6 MV (0.682) 95 5.0 99.39 0.0011916 836.0 0.992
10 MV (0.733) 3.9 99.46 0.0011905 837.1 0.993
15 MV (0.758) 7 93 4.5 95.42 0.0011347 842.4 1.000
a

Corrected to 20°C, allowing for the temperature-dependency of the specific heat capacity of graphite. A correction for the beam non-uniformity over the area of the SSC core relative to the NPL2611 chamber volume was estimated from beam profile data and found to be less than 0.1% for all energies and is not included here.