Table II.

Intra- and inter-day accuracy and precision of QCs, LLOQ and ULOQ samples in plasma.

Nominal concentration, µg/ml Intra-day a (n=5) RSD, % RE, % Inter-daya (n=5, 6 daysb) RSD, % RE, %
FTFSLDTSK
  5 4.85±0.59 12.20 −2.95 5.02±0.81 16.18 0.41
  10 9.36±1.02 10.84 −6.43 9.69±1.28 13.18 −3.09
  50 47.74±3.08 6.45 −4.52 51.43±5.92 11.51 2.86
  200 196.09±7.08 3.61 −1.95 192.79±16.41 8.51 −3.60
  400 402.44±18.24 4.53 0.61 370.10±32.57 8.80 −7.48
VLIYFTSSLHSGVPSR
  5 5.10±0.70 13.70 1.92 5.12±0.74 14.39 2.47
  10 9.70±0.34 3.48 −3.02 9.42±1.212 12.93 −5.76
  50 51.10±4.48 8.76 2.20 48.86±4.47 9.15 −2.28
  200 218.08±8.62 3.95 9.04 192.04±21.09 10.98 −3.98
  400 394.60±6.44 1.63 −1.35 375.16±27.13 7.23 −6.21
STAYLQMNSLR
  5 4.90±0.58 11.80 −2.04 5.09±0.96 18.84 1.71
  10 9.66±1.05 10.85 −3.37 10.04±1.46 14.53 0.44
  50 52.76±4.77 9.04 5.52 49.81±5.35 10.73 −0.38
  200 204.29±13.40 6.56 2.15 192.66±16.84 8.74 −3.67
  400 415.61±24.71 5.95 3.90 399.65±31.18 7.80 −0.09
a

Data are shown as mean ± SD.

b

Independent validation days for accuracy and precision. QC, quality control; LLOQ, lower limit of quantitation; ULOQ, upper limit of quantitation; RE, relative error; RSD, relative standard deviation.