Skip to main content
. 2024 Mar 27;14(4):2870–2883. doi: 10.21037/qims-23-1204

Table 3. The diameter and percent stenosis at different stenoses and reconstruction algorithms.

HIR MBIR DLR 2nd generation DLR
25% plaque
   Diameter (mm) 1.9 (0.2) 2.0 (0.2) 2.0 (0.2) 2.0 (0.1)
   Percent stenosis (%) 27.7 (8.0) 27.6 (4.3) 27.1 (4.5) 26.4 (3.8)
50% plaque
   Diameter 1.3 (0.2) 1.3 (0.2) 1.4 (0.2) 1.4 (0.1)
   Percent stenosis (%) 52.3 (9.5) 53.2 (6.8) 49.7 (6.4) 49.9 (5.2)
75% plaque
   Diameter 0.7 (0.1) 0.7 (0.1) 0.8 (0.1) 0.7 (0.1)
   Percent stenosis (%) 72.5 (5.8) 73.0 (3.8) 70.6 (4.6) 73.1 (2.9)

The diameter and percent stenosis values are presented as mean (standard deviation). HIR, hybrid iterative reconstruction; MBIR, model based iterative reconstruction; DLR, deep learning-based reconstruction.