. 2023 Oct 26;25(3):383–395. doi: 10.1093/ehjci/jead280

Figure 4.

Figure 4

Test–retest reproducibility of AI measurements compared with manual measurements. Test–retest reproducibility presented as variance ratios with bootstrap CI for AI measurements compared with reference. The horizontal lines and numbers in parentheses represent 95% CI for superiority. The upper (right) ends of the horizontal lines represent 97.5% one-sided CI for non-inferiority. Upper panel: All 95% CI were below the variance ratio of 1.0, indicating superiority of AI compared with the inter-observer scenarios. Lower panel: The upper ends of the CI were between 1.0 and the prespecified margin of 1.46, indicating non-inferiority of AI compared with the intra-observer scenarios. Abbreviations: AI, artificial intelligence; EDV, end-diastolic volume; EF, ejection fraction; ESV, end-systolic volume.