Published in final edited form as: Med Phys. 2022 May 15;49(7):4430–4444. doi: 10.1002/mp.15653

TABLE 4.

Evaluation results of the simulated scanners in GATE based on NEMA guidelines

Parameters Distancea EXPLORER Biograph Vision Prism-PET 1 mmb Prism-PET 1.5 mmb

Spatial resolution (mm) FWHM FWTM FWHM FWTM FWHM FWTM FWHM FWTM
Radial   1 2.92 5.33 3.30 6.04 1.23 2.24 1.22 2.22
  5 3.10 5.66 3.34 6.11 1.25 2.28 1.20 2.18
10 4.49 8.20 3.76 6.86 1.54 2.80 1.26 2.29
Tangential   1 2.83 5.17 3.34 6.10 1.21 2.20 1.33 2.42
  5 4.23 7.72 3.28 5.98 1.31 2.38 1.43 2.60
10 4.13 7.54 3.50 6.39 1.29 2.35 1.33 2.42
Axial   1 4.26 7.78 4.43 8.09 1.16 2.11 1.26 2.34
  5 4.30 7.85 4.47 8.16 1.19 2.17 1.31 2.30
10 5.27 9.62 4.59 8.38 1.17 2.13 1.29 2.35
Sensitivity (kcps/MBq) 197.9 16.3 50.4 52.1
Peak NECR (kcps at kBq/mL) 1367.4 at 16.8 290.0 at 25.2 923.6 at 25.2 957.8 at 25.2

Abbreviations: DOI, depth-of-interaction; FOV, field of view; FWHM, full width at half maximum; FWTM, full width at tenth maximum; GATE, Geant4 application for tomographic emission; NECR, noise equivalent count rate; NEMA, National Electrical Manufacturers Association.

a

Distance (in cm) from the center of transaxial FOV.

b

ICS + DOI corrections were applied to the Prism-PET scanners.