Table 1. MicroSight® MSI agreement to the PCR fragment analysis.

Type Extraction method PCR Instrument Laboratory MSS (agreement%) MSI-L (agreement %) MSI-H (agreement %) Cohen’s Kappa [95% CI]
Paired QIAamp MyGo Pro 1 151/151 (100) 1/1 (100) 33/33 (100) 1.000
Paired BasePurifier MyGo Pro 1 13/13 (100) 0/0 (100) 27/27 (100) 1.000
Paired BaseRelease MyGo Pro 1 13/13 (100) 0/0 (100) 27/27 (100) 1.000
Paired BasePurifier BaseTyper 1 13/13 (100) 0/0 (100) 27/27 (100) 1.000
Paired QIAamp MyGo Pro 2 25/25 (100) 0/0 (100) 5/5 (100) 1.000
Paired QIAamp MyGo Pro 3 25/25 (100) 0/0 (100) 5/5 (100) 1.000
Paired QIAamp MyGo Pro 4 25/25 (100) 0/0 (100) 5/5 (100) 1.000
Paired QIAamp MyGo Pro 5 25/25 (100) 0/0 (100) 5/5 (100) 1.000
Universal QIAamp MyGo Pro 1 151/151 (100) - 33/33 (100) 1.000
Universal BasePurifier MyGo Pro 1 13/13 (100) - 27/27 (100) 1.000
Universal BaseRelease MyGo Pro 1 12/13 (92.3) - 27/27 (100) 0.9419 [0.8295; 1.000]
Universal BasePurifier BaseTyper 1 13/13 (100) - 27/27 (100) 1.000
Universal QIAamp MyGo Pro 2 25/25 (100) - 5/5 (100) 1.000
Universal QIAamp MyGo Pro 3 25/25 (100) - 5/5 (100) 1.000
Universal QIAamp MyGo Pro 4 25/25 (100) - 5/5 (100) 1.000
Universal QIAamp MyGo Pro 5 25/25 (100) - 5/5 (100) 1.000