. 2024 Apr 11;23(2):157–163. doi: 10.7861/clinmed.2022-0252

Table 2.

Agreement between the different locations and distances of the IR temperature measurement devices and between the SpotOn and other IR temperature measurement devices

IR device Mean (SD) Mean difference (LOA)a ICC (95% CI)
IR gun
Location: eyes, nose and lips 0.15 (0.04, 0.27)
Distance head: 3, 6 and 9 cm 0.71 (0.63, 0.78)
Distance temple: 3, 6 and 9 cm 0.72 (0.65, 0.79)
Tympanic thermometer
Left and right ear 0.66 (0.54, 0.75)
SpotOn versus IR device
SpotOn 36.9 (0.37)
IR gun location
Eyes 36.9 (0.36) 0.07 (–0.88, 1.02) 0.13 (0.00, 0.30)
Nose 35.5 (1.26) 1.40 (–1.15, 3.94) 0.00 (0.00, 0.20)
Lips 36.5 (0.38) 0.41 (–0.49, 1.31) 0.25 (0.08, 0.41)
IR gun distance
Forehead: 3 cm 36.7 (0.30) 0.19 (–0.58, 0.97) 0.32 (0.14, 0.47)
Forehead: 6 cm 36.6 (0.22) 0.35 (–0.43, 1.14) 0.15 (0.00, 0.32)
Forehead: 9 cm 36.4 (0.23) 0.51 (–0.32, 1.35) 0.07 (0.00, 0.24)
Temple: 3 cm 36.7 (0.48) 0.24 (–0.93, 1.41) 0.05 (0.00, 0.22)
Temple: 6 cm 36.5 (0.42) 0.42 (–0.65, 1.49) 0.05 (0.00, 0.23)
Temple: 9 cm 36.4 (0.46) 0.57 (–0.60, 1.73) 0.00 (0.00, 0.18)
Tympanic
Left 36.3 (0.49) 0.59 (–0.37, 1.55) 0.36 (0.19, 0.51)
Right 36.3 (0.49) 0.64 (–0.31, 1.60) 0.37 (0.20, 0.51)
IR image
Coronal plane 36.4 (0.53) 0.50 (–0.58, 1.59) 0.27 (0.10, 0.43)
a

SpotOn minus comparator measurement.

CI = confidence interval; ICC = intra-class correlation coefficient; IR = infrared.