. 2024 Mar 28;14(4):190. doi: 10.3390/metabo14040190

Table 3.

Validation parameters of analytical method for crocetin quantification (results are expressed in ng eq. trans-crocetin/mL of serum).

Linearity and Sensitivity
Linear Equation Regression
Coefficient (r2)
LOD
(ng/mL)
LOQ
(ng/mL)
y = 0.241 x − 0.031 0.9988 0.09 0.3
Precision and accuracy
Nominal concentration (ng/mL) Concentration found (ng/mL) Precision (%) Accuracy (%)
Intra-day Inter-day Intra-day Inter-day
3.0 3.1 +/− 0.2 2.3 5.7 −0.5 to 13.4 6.0
16.0 15.1 +/− 0.6 3.4 4.1 −3.5 to −8.9 −5.6
29.0 27.0 +/− 0.8 2.7 3.1 −5.2 to −10.0 −7.9
Recovery and matrix effect
Recovery Matrix effect
Nominal concentration (ng/mL) Mean +/− SD RSD Nominal concentration (ng/mL) Mean +/− SD RSD
3.0 106.0 +/− 6.0 5.7 1.4 93.4 ± 2.6 2.8
16.0 94.4 +/− 3.8 4.1 14.0 96.6 ± 1.1 1.1
29.0 92.1 +/− 2.9 3.1
Stability
Conditions Spiked concentration (ng/mL) Measured concentration (ng/mL) Precision
(RSD %)
Accuracy (RE %)
Temperature (°C) Time
Room Temperature 4 h 14 13.2 ± 0.1 0.5% −8.5%
4 24 h 14.3 ± 0.8 5.6% −0.6%
−20 1 month 13.3 ± 0.5 3.6% −7.3%
3 Freeze–thaw 12.6 ± 0.3 2.0% −12.4%