Table 2.

Virus detection using Cobas VL and Xpert VL in seroconverter specimens

Virus detected using Xpert VL Virus not detected using Xpert VL Total
Virus detected using Cobas VL 144 (94.7%) 8 (5.3%)a 152
Virus not detected using Cobas VL 2 (3.3%)b 59 (96.7%) 61
a

Two had virus detected that was not in the quantifiable range (Cobas VL <1.3 log10 copies/mL or 20 copies/mL) and six had Cobas VL between 1.37 and 1.62 log10 copies/mL.

b

Both had detectable but not quantifiable Xpert VL results (i.e., <1.6 log10 copies/mL or 40 copies/mL).