Table 2.
Settings used in testing
| Settings | B125M | ePM10 | iPM12 | Life Scope | Efficia | IntelliVue |
|---|---|---|---|---|---|---|
| Default QRS threshold (mV) | ∼0.3∗ | 0.15 | 0.15 | 0.15 | 0.2 | 0.15 |
| ASY delay (s) | 5 | 5 | 5 | 5 | 5 | 4† |
| VT rate (bpm) | 100 | 100 | 100 | 100 | 100 | 100 |
| VT length (PVCs) | 6 | 6 | 6 | 9‡ | 6 | 6 |
ASY = asystole; PVC = premature ventricular contraction; VT = ventricular tachycardia.
The B125M default QRS threshold setting is approximately ∼0.3 mV, labeled on the device as “normal.”
Maximum value.
Setting unchangeable.