Table 2.

Settings used in testing

Settings B125M ePM10 iPM12 Life Scope Efficia IntelliVue
Default QRS threshold (mV) ∼0.3 0.15 0.15 0.15 0.2 0.15
ASY delay (s) 5 5 5 5 5 4
VT rate (bpm) 100 100 100 100 100 100
VT length (PVCs) 6 6 6 9 6 6

ASY = asystole; PVC = premature ventricular contraction; VT = ventricular tachycardia.

The B125M default QRS threshold setting is approximately ∼0.3 mV, labeled on the device as “normal.”

Maximum value.

Setting unchangeable.