Skip to main content
. 2024 Feb 12;5(2):70–77. doi: 10.1016/j.cvdhj.2024.02.002

Table 5.

False alarms by type

Alarm B125M ePM10 iPM12 Life Scope Efficia IntelliVue
VT 4 36 34 1 475 294
VF/VT 0 106 84 56 4 3
NSVT 11 173 164 0 0 0
AVR 3 0 0 0 0 0
Asystole 2 9 2 8 14 7
VRUN 0 0 0 227 0 0
Total 20 324 284 292 493 304

AVR = accelerated ventricular rhythm; NSVT = nonsustained ventricular tachycardia; VF = ventricular fibrillation; VRUN = ventricular beat runs; VT = ventricular tachycardia.

Labeled as VF on Life Scope.

Labeled as VT >2 on the B125M.