Skip to main content
. 2024 May 14;26(5):euae108. doi: 10.1093/europace/euae108

Table 4.

Specifications, definitions, and management strategies for patients with CIED undergoing MR procedures

Device specifications Definitions Advice for MRI
MRI-conditional All components of the device are tested as MRI-conditional Cleared for 1.5 T and if specified also for 3.0 T MRI, monitored setting advised
MRI-non-conditional Manufacturer not cleared for use in MRI conditions, combination of parts of different manufacturers 1.5 T MRI may be performed, precautions and monitoring advised
MRI-unsafe Fractured leads, lead extenders No data on MRI safety available

MRI, magnetic resonance imaging.