Aneroid Auscultatory Sphygmomanometers and Automated Oscillometric Devices as Mercury-Free Alternatives in Children

Young Hwan Song 1,
PMCID: PMC11109841  PMID: 38767341

Up to recently, measuring blood pressure (BP) with mercury sphygmomanometers has been accepted as a gold standard method for several reasons. First, it is potentially the most accurate among the currently available BP measurement methods, and second, it is the method most widely used to evaluate the risks of hypertension (HTN) and the benefits of treating HTN. In addition, mercury sphygmomanometers have only a little variation between different brands in terms of accuracy.1)

With the outbreak of Minamata disease in Japan in 1956 caused by mercury which is revealed to be a persistent bioaccumulative toxic pollutant, the international communities have become alert to the toxicity of mercury, and thus the World Health Organization (WHO) called for a phase-out of mercury sphygmomanometers from healthcare settings (https://www.who.int/publications-detail-redirect/WHO-SDE-WSH-05.08.), creating unforeseen challenges to find mercury-free alternatives to accurately measure BP, especially in pediatric population. At present, the most widely used mercury-free alternatives are aneroid auscultatory sphygmomanometers and automatic oscillometric devices.

Recently, Kim and colleagues2) found that in the comparison of measured BP values, an aneroid auscultatory sphygmomanometer correlated well with mercury sphygmomanometer, while an automatic oscillometric device did not, in Korean youths aged 10–18 years. Accordingly, they suggested aneroid auscultatory devices, not oscillometric devices, as possible alternatives for mercury sphygmomanometers in children and adolescents.

Among mercury-free alternatives, the aneroid auscultatory sphygmomanometers are most similar to the mercury sphygmomanometers, given that they directly measure systolic and diastolic BP using the auscultatory method. Thus, the recent pediatric HTN guidelines recommended that for the confirmative diagnosis of HTN, auscultatory devices should be used.1)

However, they differ from the mercury sphygmomanometers in that the mercury pressure gauge is replaced by the mechanical pressure gauge using a spring. This difference potentially causes the accuracy of aneroid devices to be low and inconsistent compared to that of the mercury sphygmomanometers. The reported data on the accuracy of aneroid devices in clinical use are quite mixed, with the frequency of defective devices varying from zero to more than 30%.3),4) In addition, the auscultatory measurement of BP requires specialized technique to perform, and thus potentially increases the probability of observer error, even though there is a lack of studies on the extent of that with aneroid devices.

The basis of the automated oscillometric method is that it directly detects mean arterial pressure, and indirectly estimates systolic and diastolic BP using specific algorithms.5) The algorithms used by different companies vary, and are never publicly disclosed. Furthermore, they may be changed from one model to another, without the user knowing. While there are many oscillometric devices that have passed the Association for the Advancement of Medical Instrumentation and British Hypertension Society protocols, these protocols evaluate monitors according to how well they perform in a population of subjects, not individual patients. Thus, it is possible that an oscillometric monitor with an average error within the recommended ± 5 mmHg tolerance for the population might have an error over 5 mmHg in some individuals.6)

Despite this, automatic oscillometric recorders have several advantages. First, it requires minimal practitioner training, and second, since it provides an automated digital reading of the BP, the observer errors are greatly reduced, especially when performing a series of readings in a short period of time, which is recommended in recent pediatric HTN guidelines to increase the reliability of the patient’s average BP, and perhaps also to minimize the white coat effect.1) In addition, it is unlikely to show the ‘drift’ over time that leads to inaccuracies of mechanical devices such as the mercury or aneroid devices. Electronic pressure transducers are usually stable over time, and if something goes wrong with the device, it is much more likely not to work at all than to give an erroneous reading.

With these advantages, oscillometric devices are particularly useful for the population screening and large-scale research studies, and have become a standard method in adults in clinical settings. Additionally, recent pediatric HTN guidelines recommended that oscillometric devices validated in children can be used for BP screening.1) Furthermore, the oscillometric method has large potential for BP measurement. For instance, this method is used almost exclusively for ambulatory BP monitoring, one of the standard tools for the diagnosis of hypertension in children and increasingly so for home BP monitoring.7),8) There are steadily accumulating evidences indicating that BP values obtained by ambulatory BP monitoring predict cardiovascular risks better than clinic measured BP. Accordingly, it is likely that the number of BP readings is more important in predicting cardiovascular risks than the accuracy of individual readings.9)

Nevertheless, in the diagnosis of HTN, the clinic BP measurement will be used persistently, and thus important clinical decisions will continue to be made on very small number of BP readings (rarely more than three). Therefore, a continuous effort to raise the accuracy of BP measuring methods is required.

Auscultatory and oscillometric devices are cuff-based sphygmomanometers. When measuring BP with these devices, patients may become uncomfortable and nervous, which may decrease the accuracy of measured BP especially in children. Recently, with the advance of information and communication technology and the widespread mobile smartphone devices and smartwatches, cuffless BP measurement devices using pulse-transit-time and photoplethysmography signals have been developed. Although they do not yet meet the usual criteria of validation and are not yet recommended by professional societies, they have great potential for BP screening and monitoring.10)

Footnotes

Funding: The author received no financial support for the research, authorship, and/or publication of this article.

Conflict of Interest: The author has no financial conflicts of interest.

Data Sharing Statement: The data generated in this study is available from the corresponding author upon reasonable request.

The contents of the report are the author’s own views and do not necessarily reflect the views of the Korean Circulation Journal.

References

  • 1.Flynn JT, Kaelber DC, Baker-Smith CM, et al. Clinical practice guideline for screening and management of high blood pressure in children and adolescents. Pediatrics. 2017;140:e20171904. doi: 10.1542/peds.2017-1904. [DOI] [PubMed] [Google Scholar]
  • 2.Kim SH, Kim YM, Kim SH, Shin JH, Lee EM. Replacing mercury sphygmomanometers with mercury-free sphygmomanometers for the National Health Survey in children: direct comparisons applying two types of mercury-free sphygmomanometer. Korean Circ J. 2024;54:270–287. doi: 10.4070/kcj.2023.0274. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Burke MJ, Towers HM, O’Malley K, Fitzgerald DJ, O’Brien ET. Sphygmomanometers in hospital and family practice: problems and recommendations. Br Med J (Clin Res Ed) 1982;285:469–471. doi: 10.1136/bmj.285.6340.469. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Canzanello VJ, Jensen PL, Schwartz GL. Are aneroid sphygmomanometers accurate in hospital and clinic settings? Arch Intern Med. 2001;161:729–731. doi: 10.1001/archinte.161.5.729. [DOI] [PubMed] [Google Scholar]
  • 5.Mauck GW, Smith CR, Geddes LA, Bourland JD. The meaning of the point of maximum oscillations in cuff pressure in the indirect measurement of blood pressure--part ii. J Biomech Eng. 1980;102:28–33. doi: 10.1115/1.3138195. [DOI] [PubMed] [Google Scholar]
  • 6.van Montfrans GA. Oscillometric blood pressure measurement: progress and problems. Blood Press Monit. 2001;6:287–290. doi: 10.1097/00126097-200112000-00004. [DOI] [PubMed] [Google Scholar]
  • 7.Urbina E, Alpert B, Flynn J, et al. Ambulatory blood pressure monitoring in children and adolescents: recommendations for standard assessment: a scientific statement from the American Heart Association Atherosclerosis, Hypertension, and Obesity in Youth Committee of the council on cardiovascular disease in the young and the council for high blood pressure research. Hypertension. 2008;52:433–451. doi: 10.1161/HYPERTENSIONAHA.108.190329. [DOI] [PubMed] [Google Scholar]
  • 8.Stergiou GS, Karpettas N, Kapoyiannis A, Stefanidis CJ, Vazeou A. Home blood pressure monitoring in children and adolescents: a systematic review. J Hypertens. 2009;27:1941–1947. doi: 10.1097/HJH.0b013e32832ea93e. [DOI] [PubMed] [Google Scholar]
  • 9.Staessen JA, Asmar R, De Buyzere M, et al. Task Force II: blood pressure measurement and cardiovascular outcome. Blood Press Monit. 2001;6:355–370. doi: 10.1097/00126097-200112000-00016. [DOI] [PubMed] [Google Scholar]
  • 10.Lee HY, Burkard T. The advent of cuffless mobile device blood pressure measurement: remaining challenges and pitfalls. Korean Circ J. 2022;52:198–204. doi: 10.4070/kcj.2021.0405. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Korean Circulation Journal are provided here courtesy of The Korean Society of Cardiology

RESOURCES