Table 2.
Steps | Measurement loops |
---|---|
Sensor cartridge calibration | – |
Equilibration | – |
Baseline measurement | 2–3 cycles: Mix 3 min, Wait 2 min, Measure 3 min |
Inject port A—oligomycin | – |
Measure | 2–3 cycles: Mix 3 min, Wait 2 min, Measure 3 min |
Inject port B—FCCP | – |
Measure | 2–3 cycles: Mix 3 min, Wait 2 min, Measure 3 min |
Inject port C—rotenone and antimycin A | – |
Measure | 2–3 cycles: Mix 3 min, Wait 2 min, Measure 3 min |