Background

Artificial intelligence (AI) is increasingly used to manage radiologists’ workloads. The impact of patient characteristics on AI performance has not been well studied.

Purpose

To understand the impact of patient characteristics (race and ethnicity, age, and breast density) on the performance of an AI algorithm interpreting negative screening digital breast tomosynthesis (DBT) examinations.

Materials and Methods

This retrospective cohort study identified negative screening DBT examinations from an academic institution from January 1, 2016, to December 31, 2019. All examinations had 2 years of follow-up without a diagnosis of atypia or breast malignancy and were therefore considered true negatives. A subset of unique patients was randomly selected to provide a broad distribution of race and ethnicity. DBT studies in this final cohort were interpreted by a U.S. Food and Drug Administration–approved AI algorithm, which generated case scores (malignancy certainty) and risk scores (1-year subsequent malignancy risk) for each mammogram. Positive examinations were classified based on vendor-provided thresholds for both scores. Multivariable logistic regression was used to understand relationships between the scores and patient characteristics.

Results

A total of 4855 patients (median age, 54 years [IQR, 46–63 years]) were included: 27% (1316 of 4855) White, 26% (1261 of 4855) Black, 28% (1351 of 4855) Asian, and 19% (927 of 4855) Hispanic patients. False-positive case scores were significantly more likely in Black patients (odds ratio [OR] = 1.5 [95% CI: 1.2, 1.8]) and less likely in Asian patients (OR = 0.7 [95% CI: 0.5, 0.9]) compared with White patients, and more likely in older patients (71–80 years; OR = 1.9 [95% CI: 1.5, 2.5]) and less likely in younger patients (41–50 years; OR = 0.6 [95% CI: 0.5, 0.7]) compared with patients aged 51–60 years. False-positive risk scores were more likely in Black patients (OR = 1.5 [95% CI: 1.0, 2.0]), patients aged 61–70 years (OR = 3.5 [95% CI: 2.4, 5.1]), and patients with extremely dense breasts (OR = 2.8 [95% CI: 1.3, 5.8]) compared with White patients, patients aged 51–60 years, and patients with fatty density breasts, respectively.

Conclusion

Patient characteristics influenced the case and risk scores of a Food and Drug Administration–approved AI algorithm analyzing negative screening DBT examinations.

© RSNA, 2024


graphic file with name radiol.232286.VA.jpg


Summary

Patient characteristics influenced the performance of a commercially available artificial intelligence algorithm analyzing true-negative screening digital breast tomosynthesis studies.

Key Results

  • ■ In a retrospective study including 4855 breast screening patients, false-positive case scores were more likely in Black patients (odds ratio [OR] = 1.5) and less likely in Asian patients (OR = 0.7) compared with White patients.

  • ■ False-positive risk scores were more likely in Black patients (OR = 1.5) and patients with extremely dense breasts (OR = 2.8) compared with White patients and patients with fatty density breasts, respectively.

  • ■ The influence of patient characteristics on algorithm performance necessitates more demographically diverse data sets for testing and training and greater transparency.

Introduction

Breast cancer screening programs reduce breast cancer mortality (1–3). Although screening mammography leads to the early detection of breast cancer, increasing radiologists’ workloads is associated with radiologist burnout, reader fatigue, and stress, ultimately leading to increased rates of interpretation errors (4–7). This may be exacerbated by the transition from digital mammography to digital breast tomosynthesis (DBT) due to the average interpretation time for DBT being twice as long (8).

There is growing interest in the use of artificial intelligence (AI) to improve the efficiency of radiologists’ work. Preliminary data suggest that AI algorithms based on deep learning applied to mammography improve radiologists’ diagnostic performance in breast cancer detection and reduce interpretation time without impeding workflow (9–13). One study (14) found that using an AI algorithm reduced false positives in interpreting screening mammography. Because breast cancer incidence and prognosis rates differ among patients of differing races and ethnicities, ages, and mammographic breast densities, determining the importance of these factors is vital to improving the ability of AI to aid radiologists’ detection of breast cancer (15–17). However, there are few demographically diverse databases for AI algorithm training, and the U.S. Food and Drug Administration does not require diverse data sets for validation (18). Furthermore, to protect their intellectual property, companies are not transparent about how their algorithms are trained.

The aim of this study was to evaluate the impact of patient characteristics (race and ethnicity, age, and breast density) on the false-positive rate of a commercially available AI algorithm when interpreting true-negative screening DBT mammograms, specifically performance in predicting cancer on the current examination (case score) and in the subsequent year (risk score).

Materials and Methods

Patient Selection

This study was Health Insurance Portability and Accountability Act–compliant and was reviewed by the institutional review board of Duke University and determined to be exempt from the need for informed consent. This retrospective cohort study identified patients with screening DBT examinations at our academic institution between January 1, 2016, and December 31, 2019, that were interpreted as Breast Imaging Reporting and Data System (BI-RADS) 1 (ie, negative) (Fig 1) (19).

Figure 1:

Patient flowchart. DBT = digital breast tomosynthesis.

Patient flowchart. DBT = digital breast tomosynthesis.

Patients with remote history of benign core biopsy were included in the initial patient cohort. Patients with prior breast surgeries (lumpectomy, mastectomy, reduction mammoplasty) or breast implant augmentation were excluded from the initial patient cohort. These patients are given a BI-RADS 2 assessment at our institution. To reduce false-positive results, BI-RADS 2 assessment screening mammograms were not included in the initial patient cohort.

The screening mammogram protocol at our institution involves DBT with synthesized two-dimensional images. All mammographic examinations during the study period were performed with a DBT mammography unit (Selenia Dimensions; Hologic). No computer-aided diagnosis, commercially available AI software, or double reading was used during the initial clinical interpretation of the screening mammograms in the cohort. All patients had 2 years of imaging follow-up without a diagnosis of atypia or malignancy.

Potential covariates were recorded: age at the time of the initial screening examination, patient-reported race and ethnicity, and mammographic breast density according to the BI-RADS Atlas from the original study interpretation. A subset of unique patients was randomly selected from the initial cohort to provide a broad distribution of four race and ethnicity groups (Fig 1): non-Hispanic White, non-Hispanic Black, non-Hispanic Asian, and Hispanic. Any patient with Hispanic ethnicity, regardless of race, was categorized as Hispanic.

There were no a priori data on the appropriate sample size for each race and ethnicity group for this type of study. Based on convenience sampling of the available patients from the initial cohort with complete race and ethnicity information, random sampling was used to select patients for each race and ethnicity group. This process continued until the patient pool for each group was exhausted or reached 1500 unique patients, whichever occurred first. This randomization process yielded the final study patient cohort.

Image Analysis

The screening DBT examinations for the final cohort were retrospectively interpreted by a commercially available AI algorithm (ProFound AI 3.0; iCAD) (20). This algorithm has been approved by the Food and Drug Administration for use within the United States to assist in detecting breast cancer in screening DBT examinations. This algorithm is based on deep convolutional neural networks and was trained offline using an enriched multivendor screening population data set collected from more than 100 institutions and including more than 6 million training images (20,21). The development and configuration of the algorithm were completed independently prior to this study. The vendor provided the use of the AI workstation, but did not have any input on study design, case selection, data analysis and interpretation, or manuscript preparation.

The algorithm provided case-level cancer detection and risk scores for screening DBT examinations. The case score ranges from 0 to 100, and per the vendor, mammograms with a score greater than 49 have findings that likely require additional workup (20). The closer the score is to 100, the more confident the algorithm is that there is a finding on the mammogram that is malignant (20,21).

The risk score is an output from a regression model without a maximum value and takes into consideration breast density and patient age for each mammogram (20,22). Per the vendor, mammograms with a risk score greater than 0.8 have the highest 1-year risk for having breast cancer on the subsequent screening mammogram (20,22). While ProFound AI offers flexibility in providing risk estimates for up to 3 years for each mammogram, only the 1-year risk estimation was used in our study because it is more accurate than the 2- and 3-year risk estimations (22).

For this study, examinations meeting these vendor-provded thresholds were defined as positive. As all screening mammogram examinations were negative and had 2 years of atypia- and cancer-free follow-up, any study classified as positive by the AI algorithm was considered a false positive. The AI algorithm relied solely on image analysis for the case score but included patient age and breast density when calculating risk score.

Statistical Analysis

All statistical analyses were performed using R (R Foundation for Statistical Computing) and SAS, version 9.4 (SAS Institute). Breast density was categorized using the four-category system defined in the BI-RADS Atlas (19), and age was analyzed by decade of life. The primary outcome of interest was the false-positive rate for the case and risk scores provided for each examination. Patient characteristics and outcomes were summarized with medians and IQRs for continuous variables and numbers of patients and percentages for categorical variables. Groups were compared using analysis of variance and the χ2 test for continuous and categorical variables, respectively. Associations between false-positive rates and patient characteristics (age, race and ethnicity, and breast density) were evaluated using multivariable logistic regression models. Correlation between case and risk scores was estimated using the Pearson correlation coefficient, and the Student t test was used to test for correlation not equal to zero. P ≤ .05 was considered to indicate statistical significance. To allow for a broader examination of potential associations in this exploratory analysis, no corrections for multiple hypothesis testing were made.

Results

Patient Characteristics

A total of 4855 unique patients were included with a BI-RADS 1 final assessment (Fig 1). There were 1316 White, 1261 Black, 1351 Asian, and 927 Hispanic patients. The median age was 54 years (IQR, 46–63 years). Most patients had either scattered fibroglandular (46% [2254 of 4855]) or heterogeneously dense (29% [1410 of 4855]) breasts. The characteristics of patients overall and by race and ethnicity are shown in Table 1. There were differences in age and breast density between race and ethnicity groups. The correlation between case and risk scores was 0.8 (95% CI: 0.8, 0.8; P < .001).

Table 1:

Patient Characteristics and Artificial Intelligence Algorithm Scores

graphic file with name radiol.232286.tbl1.jpg

Case Scores

The median case score was 20 (IQR, 9–41), with 17% (816 of 4855) of cases classified as suspicious (score > 49) by the AI algorithm; these cases represent false positives in our analysis. An example of a false-positive case is shown in Figure 2. The distribution of case scores is shown in Figure 3A, and there were statistically significant differences by race and ethnicity, age, and breast density (all P < .001) (Table 2). In multivariable analysis, there were significant differences by race and ethnicity and age (Table 3). Compared with White patients, Black patients were more likely to have suspicious case scores (odds ratio [OR] = 1.5 [95% CI: 1.2, 1.8]) and Asian patients were less likely to have suspicious case scores (OR = 0.7 [95% CI: 0.5, 0.9]). Most but not all age groups differed in the odds of false positivity compared with patients aged 51–60 years: Younger patients (aged 41–50 years) were less likely (OR = 0.6 [95% CI: 0.5, 0.7]) and older patients (aged 71–80 years) were more likely (OR = 1.9 [95% CI: 1.5, 2.5]) to have false-positive suspicious case scores, as shown in Table 3. Breast density was not associated with false positivity after controlling for age and race and ethnicity.

Figure 2:

Example mammogram assigned a false-positive case score of 96 in a 59-year-old Black patient with scattered fibroglandular breast density. (A) Left craniocaudal and (B) mediolateral oblique views demonstrate vascular calcifications in the upper outer quadrant at middle depth (box) that were singularly identified by the artificial intelligence algorithm as a suspicious finding and assigned an individual lesion score of 90. This resulted in an overall case score assigned to the mammogram of 96.

Example mammogram assigned a false-positive case score of 96 in a 59-year-old Black patient with scattered fibroglandular breast density. (A) Left craniocaudal and (B) mediolateral oblique views demonstrate vascular calcifications in the upper outer quadrant at middle depth (box) that were singularly identified by the artificial intelligence algorithm as a suspicious finding and assigned an individual lesion score of 90. This resulted in an overall case score assigned to the mammogram of 96.

Figure 3:

Graphs show final patient cohort distribution of (A) case and (B) risk scores assigned by the commercially available artificial intelligence algorithm. For each race and ethnicity group, the histogram on the left represents the number of examinations, and on the right is a visual distribution of the number of examinations. The dashed line indicates the vendor-provided threshold above which an examination was defined as positive.

Graphs show final patient cohort distribution of (A) case and (B) risk scores assigned by the commercially available artificial intelligence algorithm. For each race and ethnicity group, the histogram on the left represents the number of examinations, and on the right is a visual distribution of the number of examinations. The dashed line indicates the vendor-provided threshold above which an examination was defined as positive.

Table 2:

Case and Risk Scores by Race and Ethnicity, Age, and Breast Density

graphic file with name radiol.232286.tbl2.jpg

Table 3:

Multivariable Logistic Regression Model of Suspicious Case and Risk Scores

graphic file with name radiol.232286.tbl3.jpg

Risk Scores

The median risk score was 0.09 (IQR, 0.04–0.24), and false-positive risk scores (score > 0.8) represented 5% (240 of 4855) of cases. The distribution of risk scores is shown in Figure 3B. An example of a false-positive case is shown in Figure 4. In unadjusted analysis, there were statistically significant (all P < .001) differences in risk scores by race and ethnicity, age, and breast density, as shown in Table 2. In the multivariable analysis, there were statistically significant differences in the odds of false positivity by race and ethnicity, age, and breast density, as shown in Table 3. Specifically, older patients (OR = 3.5 [95% CI: 2.4, 5.1] for age 61–70 vs 51–60 years) and patients with denser breasts (OR = 2.8 [95% CI: 1.3, 5.8] for extremely dense vs almost entirely fatty) were more likely to have false-positive risk scores, as were Black patients (OR = 1.5 [95% CI 1.0–2.0] for Black vs White patients).

Figure 4:

Example mammogram assigned a false-positive risk score of 1.0 in a 59-year-old Hispanic patient with heterogeneously dense breasts. Bilateral reconstructed two-dimensional (A, B) craniocaudal and (C, D) mediolateral oblique views are shown. The algorithm predicted cancer within 1 year, but this individual did not develop cancer or atypia within 2 years of the mammogram.

Example mammogram assigned a false-positive risk score of 1.0 in a 59-year-old Hispanic patient with heterogeneously dense breasts. Bilateral reconstructed two-dimensional (A, B) craniocaudal and (C, D) mediolateral oblique views are shown. The algorithm predicted cancer within 1 year, but this individual did not develop cancer or atypia within 2 years of the mammogram.

Discussion

The impact of patient characteristics on artificial intelligence (AI) performance has not been well studied. We investigated the performance of a commercially available AI algorithm in interpreting negative screening digital breast tomosynthesis mammograms (Breast Imaging Reporting and Data System 1 assessment) in a demographically diverse patient cohort and found that race and ethnicity, age, and breast density influenced the odds of receiving a false-positive case score or risk score. Specifically, Black patients were more likely to have false-positive case score (OR = 1.5 [95% CI: 1.2, 1.8]; P < .001) and risk score (OR = 1.5 [95% CI: 1.0, 2.0]; P = .02) and Asian patients were less likely to have a false-positive case score (OR = 0.7 [95% CI: 0.5, 0.9]; P = .001) than White patients. In general, older patients were more likely to have a false-positive case score (patients aged 71–80 years: OR = 1.9 [95% CI: 1.5, 2.5]; P = .03) or risk score (patients aged 61–70 years: OR = 3.5 [95% CI: 2.4, 5.1]; P < .001) than patients aged 51–60 years, while patients with extremely dense breasts had more false-positive risk scores (OR = 2.8 [95% CI: 1.3, 5.8]; P < .01) than patients with fatty density breasts. Although these specific results are for one AI algorithm from one vendor, the demographically influenced performance of commercially available AI algorithms is likely widespread, given the limitations of available training and testing data sets.

The development of AI tools requires large data sets for training, testing, and validation, but there are only a limited number of data sets available, and efforts to develop demographically diverse data sets are nascent (18). The Food and Drug Administration has requirements for the development of AI tools before commercial use, but development on a demographically diverse population is not a required component (23). Furthermore, vendors do not need to be transparent about the patient characteristics of samples used in algorithm development. Our findings suggest that vendors should provide greater granularity in the performance of their algorithms in specific patient populations. The Food and Drug Administration should also develop guidelines for the demographic distribution of development data sets and greater transparency before approval.

Our study found that Black patients had 45% greater odds of false-positive case scores and 46% greater odds of false-positive risk scores than White patients. If radiologists normalize the adoption of AI recommendations based on White patients—the largest demographic group in the United States—then they risk inappropriately higher recall rates for Black patients. This has the potential to worsen health care disparities and decrease the benefits of AI assistance (24–27). However, since all AI tools are developed differently and all practices have different demographic compositions, there is no one-size-fits-all approach. Practices must understand the strengths and weakness of their AI tool in relation to their own practice environment.

Our study shows that increased breast density is associated with increased risk score and that older age is associated with both increased case and risk scores. Breast cancer incidence increases with age and is higher in premenopausal women with heterogeneously dense breasts (OR range, 1.4–1.6) and postmenopausal women with extremely dense breasts (OR range, 1.5–2.1) (28,29). Therefore, there should be an increased risk score in these patients. However, the case score is calculated only from the imaging data; therefore, the age-related association was surprising. As patients age, there are some changes to mammographic appearance, but radiologists cannot easily predict a patient’s age based on mammogram appearance (30,31). Therefore, the algorithm may be detecting some underlying changes in the image that are driving differences in performance.

The limitations of our study are as follows. This study was performed at a single institution, which may limit the generalizability of results, but this limitation was mitigated by using an enriched cohort to ensure balance in race and ethnicity distribution, as well as by using a cohort of large size to allow for subset analysis. Our false-positivity thresholds were established by the vendor, and findings may differ if another AI algorithm is used. As our study included only true-negative DBT screening mammograms and excluded cases that turned out to be positive within 2 years, the false-positive results from the AI algorithm are limited in interpretation. The variation in false-positive results between patients with different characteristics could be due to varying performance of the AI algorithm (ie, different areas under the receiver operating characteristic curve [AUC] for patients with different characteristics) or to variation in the AI algorithm operating point (ie, same AUC but operating point moving along the receiver operating characteristic curve with different characteristics).

While the commercial AI algorithm was trained on a multivendor and multi-institution data set, specifics regarding the characteristics of patients within the training data set were not available. The demographic mix of this study was purposefully chosen but does not reflect our current practice environment. To achieve this mix, we preferentially enriched some underrepresented groups, although the selection of cases within these groups was still random.

Our study included only one commercially available AI algorithm, and although demographic performance differences are likely present in other algorithms, the specifics may be different. We chose to focus on false-positive case and risk scores, but there are other outcomes of interest, including cancer detection rate, for which performance may differ based on patient characteristics. Further, the true-positive rate (sensitivity) of the algorithm was not evaluated as our cohort comprised negative screening mammograms only. The lack of comparison studies impacts our false-positive results, but does not explain our findings related to patient characteristics. The inability to compare to prior examinations is a known limitation of commercial AI software as most products do not integrate prior examinations in their analyis (32,33).

In conclusion, our study demonstrated that patient characteristics were associated with the performance of a commercially available artificial intelligence (AI) algorithm when analyzing negative screening digital breast tomosynthesis mammograms. The Food and Drug Administration should provide clear guidance on the demographic characteristics of samples used to develop algorithms, and vendors should be transparent about how their algorithms were developed. Continued efforts to train future AI algorithms on diverse data sets are needed to ensure standard performance across all patient populations.

*

D.L.N. and Y.R. contributed equally to this work.

Supported by a research grant from iCAD to Duke University. This project involves the evaluation of iCAD’s current commercial product. This research was covered under the grant and includes an iCAD coauthor. Duke authors performed the analysis independently and were not restricted in any way in reporting the results. S.M.T. supported by National Cancer Institute (grant P30CA014236).

Disclosures of conflicts of interest: D.L.N. Grant from National Institutes of Health, consulting fees from Hologic, and honorarium for lectures, presentations, speakers bureaus, manuscript writing, or educational events from the Society of Breast Imaging. Y.R. Consulting fees from iCAD; full-time employee of iCAD since June 2023; the AI algorithm used in this study was provided by iCAD. T.M.J. Support from Duke University. S.M.T. Grant to institution from National Cancer Institute. J.Y.L. Research grant to institution from iCAD; license agreement and patent pending with iCAD for MammoCAD; the computer-aided design workstation used to analyze the images was loaned by iCAD as part of research grant. L.J.G. Grants or contracts from National Institutes of Health, Department of Defense, and ECOG-ACRIN Cancer Research Group; consulting fees from Medscape and Hologic; payment or honoraria for lectures, presentations, speakers bureaus, manuscript writing, or educational events from Korean Society of Radiology; payment for expert testimony from Hancock, Daniel & Johnson; patent planned, issued, or pending (US patent no. 10,980,519); and leadership or fiduciary role for Society of Breast Imaging and National Mammography Database.

Abbreviations:

AI
artificial intelligence
BI-RADS
Breast Imaging Reporting and Data System
DBT
digital breast tomosynthesis
OR
odds ratio

References

  • 1. Cancer stat facts: female breast cancer . National Cancer Institute Surveillance Epidemiology and End Results Program . https://seer.cancer.gov/statfacts/html/breast.html. Accessed May 28, 2023 .
  • 2. Hakama M , Coleman MP , Alexe DM , Auvinen A . Cancer screening: evidence and practice in Europe 2008 . Eur J Cancer 2008. ; 44 ( 10 ): 1404 – 1413 . [DOI] [PubMed] [Google Scholar]
  • 3. Duffy SW , Tabár L , Yen AM , et al . Mammography screening reduces rates of advanced and fatal breast cancers: results in 549,091 women . Cancer 2020. ; 126 ( 13 ): 2971 – 2979 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. West CP , Dyrbye LN , Shanafelt TD . Physician burnout: contributors, consequences and solutions . J Intern Med 2018. ; 283 ( 6 ): 516 – 529 . [DOI] [PubMed] [Google Scholar]
  • 5. Stec N , Arje D , Moody AR , Krupinski EA , Tyrrell PN . A systematic review of fatigue in radiology: is it a problem? AJR Am J Roentgenol 2018. ; 210 ( 4 ): 799 – 806 . [DOI] [PubMed] [Google Scholar]
  • 6. Taylor-Phillips S , Stinton C . Fatigue in radiology: a fertile area for future research . Br J Radiol 2019. ; 92 ( 1099 ): 20190043 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Parikh JR , Sun J , Mainiero MB . What causes the most stress in breast radiology practice? A survey of members of the Society of Breast Imaging . J Breast Imaging 2021. ; 3 ( 3 ): 332 – 342 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Dang PA , Freer PE , Humphrey KL , Halpern EF , Rafferty EA . Addition of tomosynthesis to conventional digital mammography: effect on image interpretation time of screening examinations . Radiology 2014. ; 270 ( 1 ): 49 – 56 . [DOI] [PubMed] [Google Scholar]
  • 9. Pacilè S , Lopez J , Chone P , Bertinotti T , Grouin JM , Fillard P . Improving breast cancer detection accuracy of mammography with the concurrent use of an artificial intelligence tool . Radiol Artif Intell 2020. ; 2 ( 6 ): e190208 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Rodríguez-Ruiz A , Krupinski E , Mordang JJ , et al . Detection of breast cancer with mammography: effect of an artificial intelligence support system . Radiology 2019. ; 290 ( 2 ): 305 – 314 . [DOI] [PubMed] [Google Scholar]
  • 11. Shen L , Margolies LR , Rothstein JH , Fluder E , McBride R , Sieh W . Deep learning to improve breast cancer detection on screening mammography . Sci Rep 2019. ; 9 ( 1 ): 12495 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Wu N , Phang J , Park J , et al . Deep neural networks improve radiologists’ performance in breast cancer screening . IEEE Trans Med Imaging 2020. ; 39 ( 4 ): 1184 – 1194 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Raya-Povedano JL , Romero-Martín S , Elías-Cabot E , Gubern-Mérida A , Rodríguez-Ruiz A , Álvarez-Benito M . AI-based strategies to reduce workload in breast cancer screening with mammography and tomosynthesis: a retrospective evaluation . Radiology 2021. ; 300 ( 1 ): 57 – 65 . [DOI] [PubMed] [Google Scholar]
  • 14. Lång K , Dustler M , Dahlblom V , Åkesson A , Andersson I , Zackrisson S . Identifying normal mammograms in a large screening population using artificial intelligence . Eur Radiol 2021. ; 31 ( 3 ): 1687 – 1692 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Ooi SL , Martinez ME , Li CI . Disparities in breast cancer characteristics and outcomes by race/ethnicity . Breast Cancer Res Treat 2011. ; 127 ( 3 ): 729 – 738 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16. Li CI . Racial and ethnic disparities in breast cancer stage, treatment, and survival in the United States . Ethn Dis 2005. ; 15 ( 2 Suppl 2 ): S5 – S9 . [PubMed] [Google Scholar]
  • 17. Chiu SY , Duffy S , Yen AM , Tabár L , Smith RA , Chen HH . Effect of baseline breast density on breast cancer incidence, stage, mortality, and screening parameters: 25-year follow-up of a Swedish mammographic screening . Cancer Epidemiol Biomarkers Prev 2010. ; 19 ( 5 ): 1219 – 1228 . [DOI] [PubMed] [Google Scholar]
  • 18. Jeong JJ , Vey BL , Bhimireddy A , et al . The EMory BrEast imaging Dataset (EMBED): a racially diverse, granular dataset of 3.4 million screening and diagnostic mammographic images . Radiol Artif Intell 2023. ; 5 ( 1 ): e220047 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. D’Orsi CJ , Sickles EA , Mendelson EB , et al . ACR BI-RADS atlas: breast imaging reporting and data system . Reston, Va: : American College of Radiology; , 2013. . [Google Scholar]
  • 20. ProFound Breast Health Suite. iCad . https://www.icadmed.com/profoundai.html. Accessed May 30, 2023 .
  • 21. ProFound Detection. iCad . https://www.icadmed.com/breast-health/ai-breast-cancer-detection/. Accessed November 20, 2023 .
  • 22. ProFound Risk. iCad . https://www.icadmed.com/breast-health/risk-evaluation/. Accessed April 16, 2024 .
  • 23. Artificial intelligence and machine learning in software as a medical device . U.S. Food and Drug Administration . https://www.fda.gov/medical-devices/software-medical-device-samd/artificial-intelligence-and-machine-learning-software-medical-device. Updated March 15, 2024. Accessed June 22, 2023 .
  • 24. Nguyen DL , Ambinder EB , Jones MK , Hill G , Harvey SC . Improving patient comprehension of screening mammography recall lay letters . J Am Coll Radiol 2019. ; 16 ( 12 ): 1669 – 1676 . [DOI] [PubMed] [Google Scholar]
  • 25. Nguyen DL , Harvey SC , Oluyemi ET , Myers KS , Mullen LA , Ambinder EB . Impact of improved screening mammography recall lay letter readability on patient follow-up . J Am Coll Radiol 2020. ; 17 ( 11 ): 1429 – 1436 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Nguyen DL , Ambinder EB , Myers KS , Oluyemi E . Addressing disparities related to access of multimodality breast imaging services before and during the COVID-19 pandemic . Acad Radiol 2022. ; 29 ( 12 ): 1852 – 1860 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Nguyen DL , Wilson BM , Oluyemi E , et al . Disparities associated with patient adherence to BI-RADS 3 assessment follow-up recommendations for mammography and ultrasound . J Am Coll Radiol 2022. ; 19 ( 12 ): 1302 – 1309 . [DOI] [PubMed] [Google Scholar]
  • 28. Kerlikowske K , Cook AJ , Buist DS , et al . Breast cancer risk by breast density, menopause, and postmenopausal hormone therapy use . J Clin Oncol 2010. ; 28 ( 24 ): 3830 – 3837 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29. Clark GM . The biology of breast cancer in older women . J Gerontol 1992. ; 47 ( Spec No ): 19 – 23 . [PubMed] [Google Scholar]
  • 30. Ohmaru A , Maeda K , Ono H , et al . Age-related change in mammographic breast density of women without history of breast cancer over a 10-year retrospective study . PeerJ 2023. ; 11 : e14836 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31. Yildiz S , Toprak H , Aydin S , et al . The association of breast arterial calcification and metabolic syndrome . Clinics (São Paulo) 2014. ; 69 ( 12 ): 841 – 846 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Park JI , Bozkurt S , Park JW , Lee S . Evaluation of race/ethnicity-specific survival machine learning models for Hispanic and Black patients with breast cancer . BMJ Health Care Inform 2023. ; 30 ( 1 ): e100666 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33. Potnis KC , Ross JS , Aneja S , Gross CP , Richman IB . Artificial intelligence in breast cancer screening: evaluation of FDA device regulation and future recommendations . JAMA Intern Med 2022. ; 182 ( 12 ): 1306 – 1312 . [DOI] [PMC free article] [PubMed] [Google Scholar]