To the Editor:
Patients with acute respiratory distress syndrome (ARDS) often have abnormal regional pulmonary perfusion (1–3); however, the options to topographically assess perfusion in critically ill subjects are limited (4, 5). Electrical impedance tomography (EIT) is a radiation-free imaging modality that can measure perfusion in real time at the bedside (6–8). The most validated measurement method involves the first-pass kinetics analysis of an intravenous contrast solution (9). This technique requires the injection of a 10-ml bolus of 5–20% hypertonic saline through a central vein during an apnea period lasting around 30 seconds, making it suitable only for well-sedated, mechanically ventilated patients who can tolerate an extended interruption of breathing without physiological consequences. Attempts to breathe spontaneously during this extended pause usually cause failures in data acquisition, sometimes necessitating the use of paralytics.
Other methods that do not necessitate a breath hold involve monitoring pulsatile changes in lung impedance synchronized with the cardiac ejection of blood into the lung circulation. This latter technique has demonstrated a strong correlation with stroke volume but falls short in assessing regional perfusion (9). Consequently, a significant gap exists in EIT-derived regional perfusion measurement, impeding its clinical implementation and broader adoption.
As a first step to overcome the limitations of breath holds, we evaluated the ability of EIT to measure first-pass kinetics of injected contrast during uninterrupted ventilation. We compared our no-apnea method with the conventional apnea approach in both healthy and lung-injured pigs, as well as in patients with ARDS, maintaining paralysis and volume-controlled ventilation.
The study has received approval from the Massachusetts General Brigham Institutional Animal Care and Use Committee, and the Institutional Review Board, #2020P003196 (respiratory mechanics and lung perfusion assessment in mechanically ventilated patients in the operating room and in the ICU). We used eight Yorkshire pigs, weighing between 29 and 35 kg. They were anesthetized, intubated, paralyzed, and volume-controlled mechanically ventilated. EIT scans (Enlight 2100; Timpel) were taken during 30-second end-expiratory level continuous positive airway pressure and during acquisition without interrupting ventilation. The EIT machine acquired impedance, airflow, and pressure measurements at 50 Hz, through a 32-electrode belt and a proximal dedicated pressure and flow sensor. The impedance signal is composed of matrices of 32 × 32 pixels. A 10-ml bolus of 7.5% (patients) and 12% (animals) sodium chloride was injected at a steady rate of 10 ml/s using a contrast injector (Medrad ProVis) through a central venous catheter. After these initial healthy-phase acquisitions, five pigs received 3.5 ml/kg of hydrochloric acid via bronchoscopy into the lobar bronchi of each lung, to induce lung injury. After stabilization, the same procedures were repeated and compared with and without the breath hold. Two patients with ARDS were evaluated with an identical approach.
The EIT data were extracted and processed in Matlab and Labview. During the injection in the apnea phase, the hypertonic solution changes thoracic impedance as it circulates through the right heart and lungs, allowing the cardiac and pulmonary signals to be distinguished. The impedance waveform of each lung pixel is smoothed using a γ-variate function curve fitting and analyzed by calculating the fastest rate of change of the curve (9). When injected without apnea, tidal impedance changes because of ventilation interfere with the first-pass kinetics of the hypertonic flow. However, a model-based filter using impedance, airflow, and pressure signals, as illustrated in Figure 1, allows for the isolation of these cyclic changes and the recovery of the first-pass kinetics without ventilation influence, pixel by pixel (10).
Figure 1.
Patient (acute respiratory failure, male, 61 yr, body mass index 30.3 kg/m2) data showing both approaches—apnea (left) and no apnea (right)—for pulmonary perfusion estimation using saline injection in electrical impedance tomography.
The perfusion map without apnea closely resembled the apnea perfusion map at the end-expiration level (Figure 2). The pixel-by-pixel Spearman correlation between conditions with and without apnea were 0.994 (P < 0.001) in healthy pigs, 0.951 (P < 0.001) in injured pigs, and 0.957 (P < 0.001) in patients with ARDS.
Figure 2.
Representative pulmonary perfusion maps (intensity values ranging from 0 to 1) for both approaches (apnea in the first row vs. no apnea in the second row) for two patients with acute respiratory distress syndrome (first and second columns), healthy pig (third column), and injured pig (fourth column). Bland-Altman plots of perfusion values pairing the individual pixels during apnea and no apnea conditions are plotted in the third row; intensity values show the density distribution (pixel counts per bin). Average stands for (apnea + no apnea)/2 on the horizontal, and difference stands for (apnea − no apnea) on the vertical axis. Solid and dashed white lines represent bias and 95% limits of agreement, respectively.
In the study with pigs where no breath hold was required, we performed preliminary tests comparing contrast injections initiated at both inspiratory and expiratory phases. Our goal was to discern any differences in perfusion maps related to the timing of the injection. We found no discernible differences in the perfusion maps across these conditions.
We posit that our proposed method of acquiring EIT perfusion imaging without apnea offers three primary advantages:
Simplification of the procedure: Avoidance of a breath hold simplifies the measurement of regional perfusion with EIT and may facilitate its clinical implementation.
Preservation of respiratory conditions: Perfusion measurements during tidal breathing or ventilation ensure that physiological conditions remain unaltered, as the respiratory drive influences perfusion. Although our initial findings did not reveal any significant differences between breath hold and uninterrupted ventilation perfusion, this factor could be especially important for patients whose heart–lung interactions play a significant role. In such cases, the apnea abolishes the cyclical hemodynamic effect of changes in pleural pressure during ventilation. In addition, in very sick patients, even a few seconds of breath hold could cause alveoli to collapse, interfering with the perfusion distribution and leading to incorrect interpretation of data.
Expanded applicability: This approach will enable perfusion measurement in more subjects affected by cardiopulmonary conditions. Many patients who are not under paralysis cannot tolerate extended breath holds, because of respiratory drive. Any inspiratory effort during the acquisition process can compromise the first-pass kinetics analysis, whereas repeated attempts to achieve apnea to obtain a successful measurement can lead to excessive administration of sodium chloride solution. In contrast, our approach requires only an interval with a stable ventilatory pattern (20–30 s) to decompose impedance changes into ventilation and perfusion. In addition, ventilator settings could be temporarily changed to reduce respiratory drive during the examination, which is fairly feasible and safe.
Our study documents our ability to measure first-pass kinetics during uninterrupted controlled ventilation with muscle paralysis. These results support future studies to test and adapt our approach to conditions of assisted or spontaneous ventilation (e.g., assist control or pressure support ventilation) when inspiratory effort is present and the respiratory pattern is variable.
In conclusion, we aimed to address a primary challenge in estimating regional pulmonary perfusion with EIT during controlled ventilation. Our findings demonstrate that EIT signal decomposition of saline injections during steady ventilation yields assessments of regional perfusion comparable to EIT perfusion during apnea in a cohort of patients and healthy and injured pigs.
Acknowledgments
Acknowledgment
The authors thank Talisa Buehl, Gamonmas Ekkapat, Hatus Wanderley, Dario Winterton, Mohamed Okda, and all the lab members for their invaluable assistance.
Footnotes
Supported by NIH NHLBI grant R01-HL137389.
Author Contributions: M.V., Y.X., G.A., M.A., and M.C. designed the study and performed data collection and analysis. T.G., E.C., T.W., and L.B. performed data analysis and interpretation. M.V., Y.X., and M.C. drafted the manuscript. All authors contributed to a critical manuscript review and provided final approval for submission.
Originally Published in Press as DOI: 10.1164/rccm.202310-1919LE on February 27, 2024
Author disclosures are available with the text of this letter at www.atsjournals.org.
References
- 1. Perier F, Tuffet S, Maraffi T, Alcala G, Victor M, Haudebourg A-F, et al. Effect of positive end-expiratory pressure and proning on ventilation and perfusion in COVID-19 acute respiratory distress syndrome. Am J Respir Crit Care Med . 2020;202:1713–1717. doi: 10.1164/rccm.202008-3058LE. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Perier F, Tuffet S, Maraffi T, Alcala G, Victor M, Haudebourg A-F, et al. Electrical impedance tomography to titrate positive end-expiratory pressure in COVID-19 acute respiratory distress syndrome. Crit Care . 2020;24:678. doi: 10.1186/s13054-020-03414-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Spina S, Marrazzo F, Morais CCA, Victor M, Forlini C, Guarnieri M, et al. Modulation of pulmonary blood flow in patients with acute respiratory failure. Nitric Oxide . 2023;136–137:1–7. doi: 10.1016/j.niox.2023.05.001. [DOI] [PubMed] [Google Scholar]
- 4. Kay FU, Beraldo MA, Nakamura MAM, De Santis Santiago R, Torsani V, Gomes S, et al. Quantitative dual-energy computed tomography predicts regional perfusion heterogeneity in a model of acute lung injury. J Comput Assist Tomogr . 2018;42:866–872. doi: 10.1097/RCT.0000000000000815. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Kircher M, Elke G, Stender B, Hernandez Mesa M, Schuderer F, Dossel O, et al. Regional lung perfusion analysis in experimental ARDS by electrical impedance and computed tomography. IEEE Trans Med Imaging . 2021;40:251–261. doi: 10.1109/TMI.2020.3025080. [DOI] [PubMed] [Google Scholar]
- 6. Gaulton TG, Martin K, Xin Y, Victor M, Ribeiro De Santis Santiago R, Britto Passos Amato M, et al. Regional lung perfusion using different indicators in electrical impedance tomography. J Appl Physiol (1985) . 2023;135:500–507. doi: 10.1152/japplphysiol.00130.2023. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Martin KT, Xin Y, Gaulton TG, Victor M, Santiago RR, Kim T, et al. Electrical impedance tomography identifies evolution of regional perfusion in a porcine model of acute respiratory distress syndrome. Anesthesiology . 2023;139:815–826. doi: 10.1097/ALN.0000000000004731. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Frerichs I, Amato MBP, van Kaam AH, Tingay DG, Zhao Z, Grychtol B, et al. TREND study group Chest electrical impedance tomography examination, data analysis, terminology, clinical use and recommendations: consensus statement of the TRanslational EIT developmeNt stuDy group. Thorax . 2017;72:83–93. doi: 10.1136/thoraxjnl-2016-208357. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Borges JB, Suarez-Sipmann F, Bohm SH, Tusman G, Melo A, Maripuu E, et al. Regional lung perfusion estimated by electrical impedance tomography in a piglet model of lung collapse. J Appl Physiol (1985) . 2012;112:225–236. doi: 10.1152/japplphysiol.01090.2010. [DOI] [PubMed] [Google Scholar]
- 10. Victor M, Melo J, Roldán R, Nakamura M, Tucci M, Costa E, et al. Modelling approach to obtain regional respiratory mechanics using electrical impedance tomography and volume-dependent elastance model. Physiol Meas . 2019;40:045001. doi: 10.1088/1361-6579/ab144a. [DOI] [PubMed] [Google Scholar]