Table 1. Comparisons of global gamma pass rates and beam delivery times between non-gated and gated 6 MV FFF noncoplanar VMAT and IMRT deliveries using the identical patient’s CT images and dose prescriptions/constraints, where the global gamma pass rates were calculated with a threshold of 10% and a criterion of 3 mm/3% in absolute dose mode.
6 MV FFF VMAT (noncoplanar 3 arcs) | 6 MV FFF sliding-window IMRT (noncoplanar 7 ports) | |||
Non-gated | Gated | Non-gated | Gated | |
Global gamma pass rate (%) | 94.4 | 94.3 | 99.2 | 99.2 |
Beam delivery time (sec) | 133 | 373 | 140 | 375 |
Total MU | 2377 | 2270 |