Table 1. Comparisons of global gamma pass rates and beam delivery times between non-gated and gated 6 MV FFF noncoplanar VMAT and IMRT deliveries using the identical patient’s CT images and dose prescriptions/constraints, where the global gamma pass rates were calculated with a threshold of 10% and a criterion of 3 mm/3% in absolute dose mode.

The total MUs were also indicated.

FFF: flattening filter-free; VMAT: volumetric modulated arc therapy; IMRT: intensity-modulated radiotherapy; CT: computed tomography; MU: monitor unit

  6 MV FFF VMAT (noncoplanar 3 arcs) 6 MV FFF sliding-window IMRT (noncoplanar 7 ports)
Non-gated Gated Non-gated Gated
Global gamma pass rate (%) 94.4 94.3 99.2 99.2
Beam delivery time (sec) 133 373 140 375
Total MU 2377 2270