Table 4. Further Publications That Utilize SIPs and Their Associated Readout Technologies for Bacterial Detection.
Target | Application | Readout | LoD | Linear Range | Imprinting method | Reference |
---|---|---|---|---|---|---|
Klebsiella pneumoniae | Detection in hospitals | Electrochemical (CV) | 0.012 CFU/mL | 101–105 CFU/mL | Self-assembly | (120) |
Escherichia coli | Detection in milk | Thermal and Electrochemical (HTM and EIS) | 1070 CFU/mL (HTM) and 120 CFU/mL (EIS) | 102–105 CFU/mL | Microcontact imprinting | (121) |
Listeria monocytogenes | Detection in milk and pork | Optical | 103 CFU/mL | 103–105 CFU/mL | Pickering emulsion | (122) |
Staphylococcus aureus | Detection in milk and rice | Optical | 103 CFU/mL | 102–106 CFU/mL | Imprinting on magnetic particles | (123) |
Escherichia coli | Detection in water | Plate culturing | 103 CFU/mL | 103–104 CFU/mL | Bulk polymerization | (124) |