Published in final edited form as: Magn Reson Med. 2020 Oct 31;85(4):2027–2039. doi: 10.1002/mrm.28572

Table 2.

The normalized ASL signal, temporal SNR (tSNR) and cerebral blood flow (CBF) in gray matter (GM), averaged across subjects (mean ± standard deviation). Significant differences in tSNR between the labeling schemes were detected and labeled in Figure 7B. No significant difference in ASL signal or GM CBF was detected between the labeling schemes.

PCASL rect-VSI segmented-sinc-VSI sinc-VSI sBIR8-VSS sBIR-VSS-inv dual-sBIR8-VSS
Averaged signal (%) in GM 0.648 ± 0.062 0.630 ± 0.097 0.578 ± 0.102 0.620 ± 0.093 0.497 ± 0.096 0.472 ± 0.092 0.618 ± 0.111
Averaged tSNR in GM 0.989 ± 0.232 3.188 ± 0.814 2.798 ± 0.763 2.874 ± 0.772 0.926 ± 0.291 2.258 ± 0.645 1.207 ± 0.547
Averaged GM CBF (ml/100g/min) 51.5 ± 5.0 49.6 ± 7.7 55.1 ± 9.7 41.6 ± 6.3 55.6 ± 10.7 52.8 ± 10.3 56.1 ± 10.1