Table 1.
Non-B1-corrected (crude) | B1-correcteda | ||||||
---|---|---|---|---|---|---|---|
ROI method | Reader | Right (mmol/L) | Left (mmol/L) | p-value | Right (mmol/L) | Left (mmol/L) | p-value |
Large ROI mean | Reader 1a | 101.0 ± 37.4 | 92.0 ± 23.2 | 0.15 | 57.7 ± 19.7 | 53.6 ± 13.0 | 0.34 |
Reader 2a | 109.9 ± 39.5 | 100.9 ± 28.6 | 0.23 | 62.8 ± 20.1 | 58.1 ± 14.5 | 0.29 | |
Small ROI mean | Reader 1a | 137.5 ± 54.0 | 126.9 ± 35.6 | 0.25 | 77.5 ± 26.4 | 73.4 ± 17.8 | 0.52 |
Reader 2a | 137.8 ± 49.8 | 128.7 ± 30.5 | 0.41 | 78.5 ± 25.1 | 74.4 ± 16.1 | 0.56 |
The results are shown for the readers’ large and small regions of interest (ROIs); p-value denotes the statistical difference between the right- and left-sided measurements
aPhantom’s signal intensity and B1 measurement measured by Reader 1 was used to minimise the effect of differences caused by phantom ROI placement that contributes to B1-corrected measurements