. 2024 Feb 27;25(6):e14276. doi: 10.1002/acm2.14276

TABLE 1.

Action and tolerance values for M‐MLC and HD‐MLC, B 5−8 for gamma values 3−5%/1.

Action limit (%)
MLC system Measurement device Gamma criteria Tolerance limit (%) β = 5 β = 6 β = 7 β = 8
HD‐MLC Film 5%/1 mm 96 95 95 94 93
4%/1 mm 93 91 89 87 85
3%/1 mm 84 81 77 73 69
M‐MLC Film 5%/1 mm 95 95 94 93 92
4%/1 mm 90 90 89 87 85
3%/1 mm 81 80 76 72 68
HD‐MLC SRSMC 5%/1 mm 99 99 99 98 98
4%/1 mm 97 97 96 95 94
3%/1 mm 90 90 88 86 84
M‐MLC SRSMC 5%/1 mm 99 99 98 98 98
4%/1 mm 97 97 96 96 95
3%/1 mm 91 90 88 86 84