Table 2. Accuracy of the QM-qHNMR Quantitation of the Diastereomeric Composition of DFZ in Two Solvents at 600 MHza.

  acetone-d6
sample DFZ content DF
FA255 100.9 ± 0.3% 99.8 ± 0.3%
FA256 100.5 ± 0.1% 99.3 ± 0.4%
FA257 100.6 ± 0.0% 98.3 ± 0.4%
RM001 100.3 ± 0.2% 100.0 ± 0.2%
RM002 100.2 ± 0.4% 101.3 ± 0.3%
  chloroform-d
sample DFZ content DF
FA255 100.8 ± 0.8% 99.0 ± 0.2%
FA256 101.0 ± 0.5% 99.4 ± 0.1%
FA257 101.1 ± 0.3% 101.3 ± 0.5%
RM001 100.3 ± 0.2% 99.2 ± 0.1%
RM002 101.2 ± 0.3% 101.5 ± 0.0%
a

All experiments were conducted in triplicate.