Skip to main content
. 2001 Mar 3;322(7285):531–536. doi: 10.1136/bmj.322.7285.531

Table 4.

Automated blood pressure measuring devices for self measurement at the upper arm validated using the protocols of the Association for the Advancement of Medical Instrumentation and the British Hypertension Society. For the first seven devices grading criteria had not been established although the British protocol was in use.13 Devices were validated in oscillometric mode unless otherwise indicated

Device Protocol
Use Recommendation
AAMI* BHS (systolic/ diastolic)
Omron HEM-400C9 Failed Failed At rest Not recommended
Philips HP5308 (auscultatory mode)9 Failed Failed At rest Not recommended
Philips HP5306/B9 Failed Failed At rest Not recommended
Healthcheck CX-5 0600209 Failed Failed At rest Not recommended
Nissei analogue monitor (auscultatory mode)9 Failed Failed At rest Not recommended
Systema Dr MI-1509 Failed Failed At rest Not recommended
Fortec Dr MI-1009 Failed Failed At rest Not recommended
Philips HP533220 Failed C/A At rest Not recommended
Nissei DS-17520 Failed D/A At rest Not recommended
Omron HEM-705CP20 Passed B/A At rest Recommended
Omron HEM-70621 Passed B/C At rest Not recommended
Omron HEM-403C22 Failed C/C Protocol violation Not recommended
Omron HEM-703CP23 Passed NA Intra-arterial Questionable recommendation
Omron M424 Passed A/A Only abstract available; details missing Questionable recommendation
Omron MX224 Passed A/A Only abstract available; details missing Questionable recommendation
Omron HEM-722C25 NA A/A Protocol violation Questionable recommendation
Omron HEM-722C26 Passed A/A At rest in elderly people Recommended
Omron HEM-735C26 Passed B/A At rest in elderly people Recommended
Omron HEM-713C27 Passed B/B At rest Recommended
Omron HEM-737 Intellisense28 Passed B/B At rest Recommended
Visomat OZ229 Passed C/B At rest Not recommended

AAMI=Association for the Advancement of Medical Instrumentation; BHS=British Hypertension Society; NA=not applied. 

*

To meet AAMI criteria the mean difference between the device and the mercury standard must be ⩽5 mm Hg or the standard deviation must be ⩽8 mm Hg. 

To meet BHS criteria devices must achieve a grade of at least B for both systolic and diastolic measurements. Grade A denotes greatest agreement with mercury standard and D denotes least agreement.