Quantifiler™ THP DNA Standard stock diluted in Quantifiler™ THP DNA Dilution Buffer (THP) was used for the construction of the standard curve. A minimum input volume of 10 μL DNA for dilutions was used to ensure the accuracy of manual pipetting, as suggested by the Quantifiler™ Manual. The concentration ranged from 50 ng/μL (St. 1) to 0.005 ng/μL (St. 5).

Standard Concentration (ng/µL) Volume Dilution Factor
St. 1 50.000 10 µL [100 ng/µL stock] +
10 µL QuantiFiler™ THP DNA dilution buffer
2X
St. 2 5.000 10 µL [Std. 1] +
90 µL QuantiFiler™ THP DNA dilution buffer
10X
St. 3 0.500 10 µL [Std. 2] +
90 µL QuantiFiler™ THP DNA dilution buffer
10X
St. 4 0.050 10 µL [Std. 3] +
90 µL QuantiFiler™ THP DNA dilution buffer
10X
St. 5 0.005 10 µL [Std. 4] +
90 µL QuantiFiler™ THP DNA dilution buffer
10X