Skip to main content
. 2024 Jan 12;26(1):100995. doi: 10.1016/j.jocmr.2024.100995

Table 5.

Summary of vendors’ solutions for CMR of CIED patients. IR: inversion recovery; SR: saturation recovery; SAR: specific absorption rate; WIP: work in progress; CIED: cardiac implantable electronic device. *corresponds to pre-release beta versions.

Vendors LGE Cine Phase contrast T2 mapping T1 mapping Perfusion Additional safe limits
GE Wideband IR (4 kHz) 2D and 3D LGE with spoiled gradient echo, optional free-breathing and AIR Recon DL 2D and 3D spoiled gradient echo 2D and 3D gradient echo No special protocols available for CIED patients Wideband IR (4 kHz) 2D spoiled gradient echo with optional AIR Recon DL and motion correction Wideband SR (5 kHz) 2D spoiled gradient echo with optional AIR Recon DL and motion correction B1 + amplitude, whole body SAR, head SARand scan duration
Siemens Wideband IR (6 KHz) with HeartFreeze for 1.5T. WIP* for other field strengths 2D spoiled gradient-echo with optional CS acceleration 2D/4D flow with spoiled gradient-echo No special protocols available for CIED patients SASHA* with wideband saturation (4 kHz) and spoiled gradient-echo readout Wideband SR (4 kHz) with 2D spoiled gradient-echo readout “Implant Suite” WIP* for restricting B1 + rms, body SAR, head SAR
Philips Wideband IR (4 KHz) for 2D and 3D LGE WIP* 2D gradient echo protocols available 2D gradient echo protocols available No special protocols available for CIED patients No special protocols available for CIED patients No special protocols available for CIED patients “ScanWise Implant” restricts, Whole Body SAR, Head SAR, Gradient strength and slew rate
Canon Actively investigating Wideband IR solutions for 2D and 3D LGE 2D spoiled gradient echo protocols available 2D spoiled gradient echo protocols available No special protocols available for CIED patients No special protocols available for CIED patients No special protocols available for CIED patients Adaptative model for SAR control