Skip to main content
. 2024 May 31;6(3):100492. doi: 10.1016/j.ocarto.2024.100492

Table 2.

30 defects scored by three independent blinded observers on outline attachment, area coverage and hydrogel integrity after 30 ​min of knee movement on continues passive motion (CPM) machine.

1.0 ​cm2 1.5 ​cm2 2.0 ​cm2 Total ICC
Outline attachment 10/10 (100%) 8/10 (80%) 8/10 (80%) 26/30 (87%) 0.68 (95%CI 0.41–0.83)
Area coverage 10/10 (100%) 9/10 (90%) 9/10 (90%) 28/30 (93%) 0.67 (95%CI 0.39–0.83)
Hydrogel integrity 10/10 (100%) 8/10 (80%) 7/10 (70%) 25/30 (83%) 0.55 (95%CI 0.17–0.77)

Outline attachment: Circumference that is in contact with the surrounding cartilage rim. Area coverage: Total cartilage defect that is covered by scaffold. Hydrogel integrity: Hydrogel free of shape deformities, fissures or cracks. ICC: inter-observer variability of three independent blinded observers on outline attachment, area coverage and hydrogel integrity.