Skip to main content
. Author manuscript; available in PMC: 2024 Jul 8.
Published in final edited form as: Med Phys. 2023 Jun 29;50(8):e865–e903. doi: 10.1002/mp.16536

Table 7.

Monthly routine QA for EPID-based QA systems

2D PTQA
Test Technique Metric Tolerance
Dose constancy 10 × 10cm static field % Dose difference* ±2%
2D static field 25 × 25cm static field 2D γ Baseline ± 2%
Patient/test plan IMRT 2D γ Baseline ± 2%
Patient/test plan VMAT 2D γ Baseline ± 2%
Positioning 25 × 25cm static field Distance ±2 mm (non-SBRT/SRS)
@ Cardinal angles ±1 mm (SBRT/SRS)
Scaling 25 × 25cm static field Distance ±2 mm
2D Transit Dosimetry QA
Test Technique Comparison Method Tolerance
Dose constancy 25 × 25cm % Dose difference ±2%
2D static field 25 × 25cm 2D γ Baseline ± 3%
Patient/test plan IMRT 2D γ Baseline ± 3%
Patient/test plan VMAT 2D γ Baseline ± 3%
*

Mean dose of small area around the central axis no larger than 1 × 1 cm.

Baseline pass rates should be established during EPID QA system commissioning and deviations in the pass rate in either direction should be monitored. For example, if an institution’s baseline was 95% pass rate at (3%/2 mm), the tolerance limits should be 93–97% with ±2% tolerance and 92–98% with ±3% tolerance using the same criteria.