Table 2.
Chest compression effectiveness
Baseline unaided | Smartwatch-based feedback device | Defibrillator-based feedback device | |
---|---|---|---|
Depth effectiveness (%) | |||
All | 29.7 ± 27.4 | 68.4 ± 25.0† | N/A |
IH rescuers | 40.2 ± 27.4 | 75.4 ± 25.6† | 73.8 ± 37.3 |
OH rescuers | 35.9 ± 32.3 | 75.1 ± 17.4† | 81.1 ± 31.4 |
Lay rescuers | 13.0 ± 12.3 | 54.6 ± 27.4† | N/A |
Rate effectiveness (%) | |||
All | 30.1 ± 41.0 | 87.5 ± 21.4† | N/A |
IH rescuers | 27.1 ± 43.9 | 94.5 ± 5.5† | 60.7 ± 34.3‡ |
OH rescuers | 57.1 ± 44.5 | 92.8 ± 9.5† | 82.2 ± 31.9 |
Lay rescuers | 6.2 ± 7.0 | 75.2 ± 33.5† | N/A |
Data are reported as mean ± SD.
IH = in-hospital; N/A = not applicable; OH = out-of-hospital.
P < .05 between baseline unaided and smartwatch-based feedback device.
P < .05 between smartwatch-based feedback device and defibrillator-based feedback device.