Skip to main content
. 2024 Jun 7;10(12):e32551. doi: 10.1016/j.heliyon.2024.e32551

Table 9.

Summary of the validation parameters for UV–vis spectrophotometry method.

Validation Parameters Metformin HCL analysis by UV–Vis spectrophotometry
Linearity range (μg/ml) (2.50–40.00)
Correlation coefficient (r2) ≥0.99733
%Recovery (90–110 %)
Repeatability and Reproducibility % RSD <10 %