TABLE 3.

Performance of the FIA across groups of RT-PCR Ct-valuesa

Ct-value
< 25 25–30 > 30
Sensitivity 100.00 (92.29–100.00) 92.31 (66.69–99.61) 31.82 (13.87–54.87)
Specificity 88.74 (83.90–92.25) 76.86 (71.31–81.62) 73.98 (68.03–79.35)
LR+ 8.88 (6.14–12.85) 3.99 (3.04–5.24) 1.22 (0.64–2.34)
LR– 0.00 0.10 (0.02–0.66) 0.92 (0.69–1.24)
PPV 77.11 (69.96–82.97) 16.90 (9.94–27.26) 31.69 (19.54–46.98)
NPV 100 (98.09–100.00) 99.49 (97.18–99.97) 74.15 (68.74–78.91)
Accuracy 91.72 (87.76–94.73) 80.96 (75.74–85.48) 62.81 (56.72–68.61)
AUC 0.94 (0.91–0.97) 0.85 (0.80–0.89) 0.53 (0.47–0.59)
a

LR+/–: positive and negative likelihood ratio; PPV: positive predictive value; NPV: negative predictive value; AUC= area under the curve; FIA = SD Biosensor Standard F COVID-19 Ag Test.