Abstract
Significance.
This work shows the benefits of using two different magnification strategies to improve the reading ability of low vision patients using a head-mounted technology.
Purpose.
The aim of this study was to conduct a comparative clinical trial evaluating the effectiveness of two magnification strategies in a head-mounted virtual reality display.
Methods.
Eighty-eight eligible low vision subjects were randomized into two arms: 1) the full-field magnification display or 2) the virtual bioptic telescope mode. Subjects completed baseline testing and received training on how to use the device properly, then took the device home for a 2–4 week intervention period. An adaptive rating scale questionnaire (Activity Inventory) was administered before and after the intervention (home trial) period to measure the effect of the system. A simulator sickness questionnaire (SSQ) was also administered. Baseline and follow-up results were analyzed using Rasch analysis to assess overall effectiveness of each magnification mode for various functional domain categories.
Results.
Both magnification modes showed a positive effect for reading, visual information, and the overall goals functional domain categories, with only reading reaching statistical significance after correction for multiple comparisons. However, there were no significant between-group differences between the two modes. The results of the SSQ showed the magnification modes of the head-mounted display (HMD) device were overall well tolerated among low vision users.
Conclusions.
Both the full-field and virtual bioptic magnification strategies were effective in significantly improving functional vision outcomes for self-reported reading ability.
In 1992, the first head-mounted electronic low vision enhancement system, equipped with a rechargeable battery and cathode ray tube (CRT) displays, was designed to assist visually impaired patients.1,2 Considering the rapid growth of low vision prevalence rates,3 the continued development of these head-mounted electronic technologies into more visually comfortable and affordable displays remains a priority to address the major functional difficulties of individuals with visual disabilities. Current designs that use smartphones as virtual reality head-mounted displays have improved rapidly and become more mainstream. These virtual reality systems have been adapted to magnify environmental images, faces, and text for those with vision impairment.4–6
We revisit concepts of magnification to understand the major limitations of current head-mounted display designs. There are three ways to magnify an image: 1) physically move closer to the object to provide relative distance magnification, 2) physically make the object larger with relative size magnification, and 3) view the object through an optical system which provides angular magnification. The first two ways - relative distance and relative size magnification - are considered linear magnification, because they simply rescale one side of the triangle that defines the visual angle of the object on the retina. In contrast, in angular magnification, an optical system increases the visual angle subtended by the object, without changing the size or distance of that object. Two examples of angular magnification include viewing an object through a telescope or viewing a magnified image using a smartphone camera (pixel magnification). In both instances, the optical system magnifies, by producing an intermediate real or virtual image that appears magnified, without changing the dimensions or distance of the viewed object. These distinctions in magnification type are important to make, as we consider some of the undesirable effects of angular magnification used in current electronic low vision enhancement system designs.
One such drawback in current head-mounted display designs, namely the algorithms used to produce the digital image transformations, is the failure of our visually enhanced visual-vestibulo-ocular reflex to compensate for magnified image motion. To understand visual-vestibulo-ocular reflex, it is important to define the vesibulo-ocular reflex or vestibulo-ocular reflex. In normal head rotation, the gain of the vestibulo-ocular reflex is defined as the ratio of the compensatory eye velocity in response to the opposing head velocity. If the system is intact, the vestibulo-ocular reflex gain is ideally 1.0 and retinal image slip does not occur. However, when there is a magnified viewing system (i.e., telescope, head-mounted display) involved, the visual-vestibulo-ocular reflex reflex is at play. Thus, the compensatory visual-vestibulo-ocular reflex gain must match the magnification factor of the viewing system. If not, retinal image slip will occur causing motion sickness.7,8 Demer has shown that retinal image stability increases the likelihood low vision patients will tolerate the use of a telescopic aid in real world environments.8 There are two reasons for this: 1) dynamic or functional visual acuity improves with head (and retinal image) stabilization, and 2) improved retinal image stability reduces vestibular ocular mismatch responsible for motion sickness. 8 Thus, strategies to improve retinal image stabilization through a magnified viewing system is key for minimizing adverse symptoms and for improving overall performance with the device.
Current designs have improved image processing speed but there remains room for improvement to minimize the effect of head or camera movements while using an electronic low vision enhancement system. The head-mounted display system is slow to process and display what is captured, and this lag in processing may contribute to the visual discrepancy which causes adverse symptoms. Moreover, our knowledge about the prevalence of visual discomfort symptoms when low vision subjects use head-mounted display devices is limited.5
Knowing these limitations, we compared two magnification strategies using head-mounted displays, to deal with this magnified image motion9: 1) a full-field magnification view and 2) a virtual bioptic telescope display. In the full-field magnification view, the user can magnify an object on a virtual projection screen using a touchpad on the head-mounted display device. We refer to the first display mode as full-field magnification as the user can magnify an image within the full view displayed on the head-mounted display device. The rationale for the second design, the virtual bioptic telescope, was to employ an unmagnified surround image that can be used to minimize the negative effects of magnified image motion. The virtual bioptic telescope mode resembles a display with an adjustable central bubble embedded within a large unmagnified field of view. The magnification within the bubble (or virtual bioptic telescope) could be varied by the user using a touchpad on the headset. In this way, the bubble provides a stable reference to the unmagnified surrounding image. Since all of the commercially available head-mounted displays magnify the entire region in the display, we test the value of using a virtual bioptic telescope system, with this unique bubble display, as a mechanism to minimize any observed discrepancy with head or camera movements.
Both magnification strategies included a virtual reality projection screen. With this feature, snapshots were taken with the Samsung hardware and texture-mapped onto the virtual projection screen. These static images could be adjusted to a desired magnification level, allowing for the complete elimination of accelerated image motion with natural head movements. Thus, as you move your head, the system pans the virtual screen in the opposite direction to compensate for any head movements, to make it appear like the environment is still, and to ultimately minimize artifactual image motion caused by head motion. This form of virtual digital image processing, including texture mapping, is further described by Deemer et al.9 Additional features of the head-mounted display hardware included reverse polarity contrast and automatic gain control to maintain constant display luminance with ambient illumination changes.
As these systems are developed and become increasingly available to patients in clinical vision rehabilitation settings, it is important to evaluate these parameters in the display to optimize the magnification capabilities and to maximize effectiveness and visual comfort with respect to completing specific visual tasks10 (e.g., reading continuous text, visual information/face identification, visual motor skills, and mobility). In a preliminary observational study,9 Deemer et al. found a significant improvement in functional vision outcomes related to visual information and reading tasks, while using both novel methods of magnification described. We extend this work by reporting our findings in this randomized comparative effectiveness trial, which evaluated the effects of these two magnification strategies: 1) the full-field magnification display and 2) the virtual bioptic telescope mode.
METHODS
Participants
We assessed 157 patients from the Johns Hopkins Wilmer Eye Institute Low Vision Rehabilitation Service for eligibility. Eighty-eight patients were enrolled in the study after meeting the following inclusion criteria: 1) best corrected visual acuity (using Early Treatment Diabetic Retinopathy Study chart) between 0.48 logMAR (20/60) and 1.6 logMAR (20/800), 2) horizontal visual field greater than 70° diameter, 3) vertical visual field greater than 50° diameter, and 4) at least 14 years old. The acuity cutoff was determined given the display resolution of the headset (3 arcmin/pixel); visual field cutoffs were selected based on the maximum fields of view displayed by the two magnification modes. Goldmann visual fields were used to determine the minimum horizontal and vertical visual field limits of participants. We excluded any patients who had a planned ocular surgery (including laser surgery) during the study, were early in anti-neovascular treatment (<6 injections), had less than 2 goals identified on the Activity Inventory (AI) Visual Function Questionnaire,10 or received a score less than 27 on the Telephone Interview for Cognitive Status (TICS).11
The study protocol was approved by the Institutional Review Board of Johns Hopkins University School of Medicine. All participants gave informed consent for participation after discussing the nature, risks, and benefits of the study.
Study Design
Participants were randomized into two arms (n=44 full-field, n=44 virtual bioptic) with the following head-mounted display specifications: 1) with full-screen magnification in a 35° field of view, and 2) with a magnification virtual bioptic embedded in a large unmagnified 70° field of view. Participants completed baseline testing then took the device home for a 2–4 week intervention period. Only those patients who demonstrated proficiency in the operation of the device after training were eligible to participate in the study. Participants were encouraged to use the device in their daily activities (excluding mobility). The trial was registered on clinicaltrials.gov as NCT03728660.
Randomization and Masking
The magnification condition was randomized by a Research Coordinator (RC) prior to the subject’s first study visit using a randomization table. A Research Assistant (RA) collected the data and was kept masked to the assigned magnification modes throughout the study for all participants. Subjects were instructed not to describe the scenes viewed through the device to the RA to keep the RA masked throughout the trial.
After the RC determined which condition the patient would have, they set up the software for each headset specific to the patient, and a separate device trainer (low vision therapist) reviewed how to use the device properly with each subject prior to the subjects taking the device home for the intervention period. Only the RC and device trainer knew the assigned magnification condition for each subject. The PI and all other team members were kept masked to the magnification modes assigned to subjects. All headsets were indistinguishable from each other; settings could be changed by only the device trainer to either of the two magnification modes.
Head-mounted Display Device
Low vision subjects were fitted with a Samsung Gear head-mounted display that used a Galaxy S8 smartphone which was inserted in the virtual reality head-mounted display (Figure 3). Custom software was developed so the smartphone screen displayed duplicate split images to each eye separately using a biocular system. Within the headset, to maintain a single comfortable image, subjects could adjust the lateral separations of the images seen through the viewing system relative to the fixed optical centers on the head-mounted display display. The smartphone camera displayed the same live video images of the environment to both eyes (with no binocular disparity) on the S8 OLED (organic light-emitting diode) display. The user could adjust the size of the virtual bioptic telescope (bubble) and the amount of magnification, with a touch pad on the side of the head-mounted display. Contrast enhancement could also be switched on and off using a button above the touch pad. The same head-mounted display hardware was used for both arms of the study with the variable features modified through the software of the device. Usage data was tracked and recorded by a built-in sensor when the device was headborne. A supplemental video was prepared to demonstrate the basic functions of how to operate the device including the swiping function used on the touchpad to magnify.
Figure 3.

Samsung Gear HMD (head-mounted display) device front view (left) and side view (right).
Device Training
Prior to taking the head-mounted display device home, participants spent at least 30 minutes with an experienced certified low vision therapist, who trained each participant about how to use the device properly (e.g., turning device on/off, using the touchpad to magnify, using the contrast enhancement feature). In addition to this training, participants were instructed not to expose the device to water and to use the device only while seated (not ambulating or driving). They were encouraged to use the device to improve their vision for any distance or near tasks during the 2–4 week intervention period. Depending on which arm of the study the participant was randomized to, the proper magnification mode was turned on or off by the low vision therapist, to display either the full-field magnification or virtual bioptic telescope mode on the head-mounted display.
Assessment of Outcomes
The Activity Inventory10 is an adaptive rating scale questionnaire that provides functional ability estimates based on difficulty ratings reported by the patient involving a wide range of daily activities. The AI consists of 510 calibrated items in the item bank and are categorized into an overall goal level (i.e., cooking a meal, managing finances) and various task levels nested under the goals, including reading, mobility, visual information (seeing/information gathering category), and visual motor functions (involving eye hand coordination), which have mutually exclusive items.10,12 We administered the AI questionnaire at the baseline visit before the home trial and at the follow-up visit (after the home trial) to measure the relative effectiveness of the intervention (full-field magnification vs. virtual bioptic telescope) on self-reported functional ability measures.
We also administered an adapted version of the Simulator Sickness Questionnaire (SSQ)13,14 twice by phone during and after the home trial to assess whether any participants reported adverse side effects while using the head-mounted display device. We developed an adapted version of this SSQ questionnaire using the original tool derived to assess visual discomfort among flight simulator participants,13 in addition to incorporating questions used on the Convergence Insufficiency Treatment Trial Symptom Survey.14 The survey included 19 questions that asked participants to rate their symptoms of physical and/or visual discomfort (e.g., eye tiredness, eyestrain, neck/back strain, or nausea) while using the device (Appendix). Normative data from a large data set of normally sighted video display users were used to calibrate the individual item measures of the SSQ survey so user responses could be interpreted.15
Statistical Analysis
To resolve a medium effect (Cohen d=0.5) with a power of 0.8, an alpha of 0.05, and a 1:1 allocation ratio, we estimated a total sample size of 100. However, data collection was discontinued for 6 months due to COVID which limited the final number of participants. The primary outcome measure for the comparative trial was the mean change score between groups in patients’ self-reported functional ability estimates, as assessed by the AI. Person measure (ability estimates) change scores were calculated using Rasch analysis.16 Cohen’s d was calculated to evaluate within group and between group effect sizes at the goal level and for reading, visual information, visual motor, and mobility tasks. Two tailed t-tests using 95% confidence intervals were used to look at between group differences. A minimum clinically important difference (MCID) for each visual ability measure was estimated as a clinical endpoint for each participant by subtracting the baseline visual function measure from the follow-up visual function measure and dividing by 1.96 times the corresponding standard error of the baseline visual function measure estimate (i.e., MCID is measured as the change >95% confidence limit on the baseline measure).
The secondary outcome of the study was between group differences in person measure scores on the SSQ. All AI and SSQ baseline and follow-up person measures were estimated on an interval scale from a Rasch model17 using the R package.
RESULTS
Participants
Eighty-eight patients (n=44 full-field, n=44 virtual bioptic) were initially enrolled in the study. One participant in the full-field group did not complete the study because at the time of the baseline visit the device was deemed unsuitable for the participant to use properly after multiple training sessions. Five participants dropped out from the full-field group and 6 dropped out from the virtual bioptic group. These participants did not want to participate anymore and opted to discontinue the study. Thus, 76 participants completed the study. The sample size of 100 was not reached as data collection was discontinued 6 months due to COVID shutting down human research data collection activities.
There was no significant difference in age (p=0.75), best-corrected visual acuity (p=0.078), contrast sensitivity (p=0.071), or baseline AI person measures between those who dropped out and those who completed the study.
Baseline characteristics of study participants are shown in Table 2. Participants represented a range of ocular diagnoses; 23% retinal dystrophy, 11% age-related macular degeneration, 14% glaucoma, 7% proliferative diabetic retinopathy, 17% other macular disease, 13% other optic nerve disease, and 16% other (including anterior segment conditions). Data was analyzed for 76 participants (n=38 full-field, n=38 virtual bioptic). Of the 76 participants who completed the study, 32 were female and 44 were male. The mean age for all subjects was 59 (age range 15–89). The mean age for the full-field group was 62 and 57 for the virtual bioptic group (p = 0.23). There was no difference in age, best corrected visual acuity, contrast sensitivity, or baseline AI person measures between groups. Visual acuity and contrast were similar between groups as detailed in Table 2.
Table 2.
Baseline characteristics of study participants.
| Characteristic | Full-field (n=38) | Virtual Bioptic (n=38) |
|---|---|---|
| Mean age (SD) | 62 (19.30) | 57 (19.18) |
| Sex | 22 males, 16 females | 22 males, 16 females |
| Best corrected visual acuity in better seeing eye, mean (SD), logMAR | 0.88 (0.25) | 0.85 (0.28) |
| Contrast sensitivity, mean (SD), log units | 1.01 (0.38) | 1.11 (0.42) |
Table 3 shows the person measures from AI broken down by visual function domain (Goals, Reading, Visual Information, Visual Motor, and Mobility) for the two magnification groups. Rasch analysis was used to calculate person measure scores; values are expressed in logits.
Table 3.
Activity Inventory person measure scores measured at baseline and after device home trial.
| AI person measure (SD), logits | Full-field (n=38) | Virtual Bioptic (n=38) |
|---|---|---|
| Baseline Mean - Goals | 1.036 (0.69) | 1.316 (1.20) |
| Mean Change Score - Goals | 0.256 (0.65) | 0.293 (0.94) |
| Baseline Mean - Reading | −2.618 (0.67) | −2.502 (1.19) |
| Mean Change Score - Reading | 1.031 (1.04) | 1.172 (1.43) |
| Baseline Mean – Visual Information | 0.250 (0.88) | 0.238 (0.89) |
| Mean Change Score – Visual Information | 0.302 (0.82) | 0.439 (1.09) |
| Baseline Mean – Visual Motor | 0.461 (1.07) | 0.573 (0.93) |
| Mean Change Score – Visual Motor | −0.189 (1.36) | −0.017 (1.31) |
| Baseline Mean - Mobility | 0.141 (0.91) | 0.564 (1.35) |
| Mean Change Score - Mobility | −0.056 (1.04) | −0.341 (1.61) |
Primary Outcome
Table 4 shows effect sizes, from baseline to follow-up, for the two groups by functional domain in the 5 categories: 1) Overall Goals, 2) Reading, 3) Mobility, 4) Visual Information, and 5) Visual Motor.
Table 4.
Within group and between group Cohen’s d effect size by functional domain for two magnification modes. (*) indicates significance equivalent to the 0.05 level after correcting for multiple comparisons (p < 0.01).
| Cohen’s Effect Size by Functional Domain | |||||
|---|---|---|---|---|---|
|
| |||||
| Goals | Reading | Mobility | Visual Info | Visual Motor | |
| Group 1 – Full Field | 0.394 (p = 0.02) |
0.995* (p = 4.15×10−7) |
−0.053 (p = 0.74) |
0.369 (p = 0.03) |
−0.139 (p = 0.40) |
| Group 2 – Virtual Bioptic Telescope | 0.313 (p = 0.07) |
0.822* (p = 1.15×10−5) |
−0.211 (p = 0.21) |
0.401 (p = 0.02) |
−0.013 (p = 0.94) |
| p-value (2-tailed between group t-test) | 0.221 | 0.624 | 0.368 | 0.540 | 0.581 |
Within Groups 1 and 2, we see a small effect (0.20 ≤d≤ 0.50) for Goals (full-field d=0.394, virtual bioptic d=0.313) and Visual Information (full-field d=0.369, virtual bioptic d=0.401), and a large effect (d≥0.80) for Reading (full-field d=0.995; virtual bioptic d=0.822). Neither magnification modes improved function for Mobility (full-field d=−0.053; virtual bioptic d=−0.211) or Visual Motor categories (full-field d=−0.139; virtual bioptic d=−0.013). The mean virtual bioptic size used by subjects in Group 2 was 30.6° ± 10.8° (SD).
When we look at between group differences, p-values indicated no significant differences between the two magnification modes for all functional categories.
We also calculated the minimal clinically important differences between the two groups in Table 5. These are the % patients in each group that had a difference in person measures which exceeded the 95% confidence interval.18 The minimal clinically important differences frequencies show relatively higher rates of improvement for Reading (full-field=50%, virtual bioptic=47%) and Visual Information (full-field 37%, virtual bioptic=39%), compared to Visual Motor, Mobility, and Goals categories. Thus, nearly half of the subjects in both magnification groups exceeded the 95% confidence interval for a significant positive effect for reading ability, in addition to more than a third of subjects exceeding the MCID threshold for visual information tasks. Calculated odds ratios were not significantly different from 1 for any of the five functional domain categories.
Table 5.
Comparing minimal clinically important difference (%) and odds ratios for each group by functional domain.
| MCID Frequency | |||||
|---|---|---|---|---|---|
|
| |||||
| Goals | Reading | Mobility | Visual Info | Visual Motor | |
| Group 1 – Full-field | 16% | 50% | 18% | 34% | 24% |
| Group 2 – Virtual Bioptic Telescope | 18% | 47% | 16% | 42% | 22% |
| Odds Ratio | 0.83 | 1.11 | 1.17 | 0.71 | 1.12 |
MCID = minimal clinically important difference
Secondary Outcome (SSQ)
To summarize some of the significant responses on the SSQ, 11% (n=4 in Group 1, n=4 in Group 2) reported headache and 4% of participants (n=2 in Group 1, n=1 in Group 2) reported symptoms of nausea while using the system. 33% (n=13 in Group 1, n=12 in Group 3) reported experiencing eye strain with 21% (n=9 in Group 1, n=7 in Group 2) classifying it as “barely,” 9% (n=3 in Group 1, n=4 in Group 2) as “moderately,” and 3% (n=1 in Group 1, n=1 in Group 2) as “very.”
Figure 5 shows the anchored SSQ items measures with person measures categorized by group. Higher values on the interval SSQ symptom severity scale indicate more severe symptoms. Zero on the scale signified the average symptom severity for the calibration of samples that are described by the SSQ items. There was a small percentage of patients whose person measure scores fell within the item measure distribution. The minimal overlap between the calibrated item measures (green) and the person measures (red/gray) signifies the small percentage of individuals who report significant negative symptoms while using the device. The majority of individuals fell on the end of the scale with minor to no symptoms. The mean SSQ person measure for the full-field and virtual bioptic groups were −4.23 and −3.99 respectively. The difference between the groups was not statistically significant.
Figure 5.

Wright map histograms of Simulator Sickness Questionnaire (SSQ) person measures for each group and anchored item measures. Higher values on the interval SSQ symptom severity scale indicate more severe symptoms. There was a small percentage of patients whose person measure scores fell within the item measure distribution, indicating that the majority of individuals fell on the end of the scale with minor to no symptoms.
The median time for overall usage with the head-mounted display device in the full-field group was 8.29 minutes/day (IQR 3.13–23.48 minutes/day) and 9.71 minutes/day (IQR 5.53–24.58 minutes/day) for the virtual bioptic group. As stated in the introduction, the virtual reality projection screen or snapshot mode was incorporated as a feature in both displays, to minimize image motion experienced with high levels of magnification. The median rate of use for this snapshot mode in the full-field group was 0.11 minutes/day (IQR = 0 – 0.55 minutes/day), and 0.02 minutes/day (IQR = 0 – 0.26 minutes/day) in the virtual bioptic group.
DISCUSSION
In this trial, we compared the effectiveness of two magnification modes for an head-mounted display device: 1) full-field magnification versus 2) virtual bioptic telescope. Both magnification modes showed a positive effect for reading, visual information, and the overall goals functional domain categories, with only reading reaching statistical significance after correction for multiple comparisons. There were no significant between-group differences when we compared the two types of magnification modes. The results of the SSQ showed the magnification modes using the head-mounted display device were overall well tolerated in majority of the participants.
As expected, we did not see any positive effects for the mobility and visual motor categories. This was no surprise as we instructed our participants to avoid ambulating or driving with the device for safety reasons. In addition, given the optical limitations of the system lacking binocular disparity, we knew that subjects would have difficulty with visual motor tasks requiring eye hand coordination and stereopsis.
Since head-stabilizing strategies improve performance with telescopic viewing systems,8 we wanted to investigate a magnification display that would improve the overall viewing experience by minimizing symptoms of visual discomfort caused by magnified image motion. One way to get rid of the magnifying image motion is with the virtual bioptic telescope display. The virtual bioptic bubble contains the region of interest, and as the user moves their head (while wearing the device), that region of interest stays within the virtual bioptic bubble. The size of the bubble and magnification could be adjusted by the user. The surrounding image outside the bubble remains unmagnified, serving as a reference to ultimately minimize the disorienting effects of the gain control that the user is unable to compensate for while making head movements. Thus, when the user changes fixation to another object, or makes a saccade, the eye moves within the stable bubble display, which ultimately minimizes the perception of magnified image motion. Nonetheless, when we compare the severity of symptoms using the revised SSQ survey, there was no significant difference seen with symptoms between the two magnification modes (Figure 5).
We also analyzed usage data to investigate how often users were using the snapshot mode. In the snapshot mode, static images could be adjusted to a desired magnification level, allowing for the complete elimination of this accelerated image motion with natural head movements. This feature was available for both magnification groups. To investigate whether this snapshot mode may have contributed to the low symptoms reported by users in both groups, we looked at median rates of usage for both groups which were low: 0.11 minutes/day (IQR = 0 – 0.55 minutes/day) for the full-field group and 0.02 minutes/day (IQR = 0 – 0.26 minutes/day) for the virtual bioptic group. Thus, the overall low usage rate for this feature likely does not explain the low symptoms reported in both groups using the SSQ. Similarly, in a recent study, Luo found that in his cohort of 16,000 smartphone magnifier users, the live image mode was used more often than the static snapshot mode described.19
When we tested the functional impact of this magnification strategy, our findings were consistent with Deemer’s preliminary observational study,5 which also showed a significant improvement in functional vision outcomes related to reading tasks, while using both full-field magnification and virtual bioptic displays. In Deemer’s study, however, the participants were experienced users of conventional low vision devices. This may be one reason for the higher effect sizes achieved in her study, as those participants were already skilled with low vision technology and may have been better equipped with using an head-mounted display device. In a separate multicenter prospective trial with 51 participants using a separate head-mounted display device, similar functional improvements were observed in overall visual ability, reading, and visual information tasks.6 However, Costela et al. found a deterioration in performance for video comprehension using a similar virtual bioptic display for people with good vision.20
We also note that the mean virtual bioptic size used by subjects in Group 2 was 30.6ᵒ, compared to the 40ᵒ full-field magnification view in Group 1. The calculated mean bubble size showed the magnified region encompassed less than half of the overall 70ᵒ field of view, confirming Group 1 subjects generally did not use the bubble as a full-field magnifier. Preference for virtual bioptic size will need to be investigated further as it relates to visual acuity loss and/or scotoma type.
Since little is known about the prevalence of visual discomfort symptoms when using these head-mounted display devices among low vision subjects,5,9 we used our adapted SSQ questionnaire to see whether 1) a high percentage of head-mounted display users in the trial had negative symptoms and 2) if there was a difference in symptoms between the two magnification groups. Our results showed low symptom severity scores for each group with no significant differences between the groups. These findings were comparable to Deemer’s observational study using a similar head-mounted display device.9 We also found that one study reported an odds ratio of 0.12 of low vision participants experiencing headaches while using a different head-mounted display device, which was similar to our odds ratio of 0.19 for the 11% participants reporting headaches in this study.21
One major limitation of the study was that we ran out of time and resources for data collection activities because of COVID 19 shutdown. This limited the final number of study participants which was the major reason for not achieving the planned sample size. In addition, we used self-reported difficulty ratings (Activity Inventory) to assess the primary outcome by determining the effect sizes from baseline to follow-up for the various functional domains. Future investigations using performance measures would further add to the results of this trial.
The results of our study have demonstrated the promise in a novel display for magnifying a region of interest in a virtual bioptic telescope design. In our investigation, this strategy did not turn out to be significantly more helpful compared to a full-field magnification display. We anticipate the advancement and development of head-mounted display technology adapted for the use of visually impaired subjects using better accelerometers, eye trackers, and gyroscopes to enhance the overall viewing experience for our patients.5 There remains room for improvement in how we can integrate these systems into a more comfortable display, so we can more accurately compensate for the linear and angular motion involved with head and eye movements of using these wearable technologies.
Supplementary Material
Figure 1.

Consort flow diagram outlining the research protocol.
Figure 2.

Flow diagram of research protocol. ETDRS = Early Treatment Diabetic Retinopathy Study; AI = Activity Inventory; TICS = Telephone Interview for Cognitive Status; VA = visual acuity; VF = visual field; h = horizontal; v = vertical; SSQ = Simulator Sickness Questionnaire.
Figure 4.

Comparison of virtual bioptic telescope mode using the embedded magnification bubble (left) and Full-field magnification mode (right).
Table 1.
Description of two types of magnification strategies used.
| Group 1 – Full-Field | Group 2 – Virtual Bioptic Telescope |
|---|---|
| Magnification display with 40° horizontal by 30° vertical field of view | 70° horizontal by 50° vertical field of view with adjustable magnification virtual bioptic (bubble) embedded in the view |
| Virtual projection screen/snapshot | Virtual projection screen/snapshot |
ACKNOWLEDGMENTS
We wish to thank our study coordinator Paul Kilmon for his efforts on this project.
Footnotes
SUPPLEMENTAL DIGITAL CONTENT
The supplemental video demonstrates the basic functions of how to operate the head-mounted display device used in the study. It shows how participants put the device on, how to focus the device, how to adjust the various modes, and how to use the touchpad/buttons to magnify on the device.
The appendix contains a copy of the adapted version of the Simulator Sickness Questionnaire (SSQ) questionnaire used in the study to assess whether any participants reported adverse side effects while using the head-mounted display device.
Contributor Information
Robert Chun, State University of New York, College of Optometry, New York, New York.
Ashley Deemer, Marshall B. Ketchum University, Southern California College of Optometry, Fullerton, California.
Kyoko Fujiwara, Wilmer Eye Institute, Johns Hopkins University School of Medicine, Baltimore, Maryland.
James Deremeik, Wilmer Eye Institute, Johns Hopkins University School of Medicine, Baltimore, Maryland.
Christopher K. Bradley, Wilmer Eye Institute, Johns Hopkins University School of Medicine, Baltimore, Maryland.
Robert W. Massof, Wilmer Eye Institute, Johns Hopkins University School of Medicine, Baltimore, Maryland.
Frank S. Werblin, University of California, Berkeley, Berkeley, California.
REFERENCES
- 1.Massof RW, Rickman DL, Lalle PA. Low-vision enhancement system. J Hopkins Apl Tech D 1994;15:120–5. [Google Scholar]
- 2.Massof RW, Rickman DL. Obstacles encountered in the development of the low vision enhancement system. Optom Vis Sci 1992;69:32–41. [DOI] [PubMed] [Google Scholar]
- 3.Varma R, Vajaranant TS, Burkemper B, et al. visual impairment and blindness in adults in the United States. JAMA Ophthalmol 2016;134:802. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.LWW MOVE TO TEXT AS: Mehta M, et al. IOVS 2018;59:ARVO E-Abstract 633.
- 5.Deemer AD, Bradley CK, Ross NC, et al. Low vision enhancement with head-mounted video display systems: Are we there yet? Optom Vis Sci 2018;95:694–703. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Wittich W, Lorenzini MC, Markowitz SN, et al. The effect of a head-mounted low vision device on visual function. Optom Vis Sci 2018;95:774–84. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Demer J, Porter F, Goldberg J, et al. Dynamic visual acuity with telescopic spectacles. Invest Ophthalmol Vis Sci 1988;29:1184–9. [PubMed] [Google Scholar]
- 8.Demer J, Porter F, Goldberg J, et al. Predictors of functional success in telescopic spectacle use by low vision patients. Invest Ophthalmol Vis Sci 1989;30:1652–65. [PubMed] [Google Scholar]
- 9.Deemer AD, Swenor BK, Fujiwara K, et al. Preliminary evaluation of two digital image processing strategies for head-mounted magnification for low vision patients. Transl Vis Sci Technol 2019;8:23. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Massof RW, Ahmadian L, Grover LL, et al. The Activity Inventory: An adaptive visual function questionnaire. Optom Vis Sci 2007;84:763–74. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Brandt J, Spencer M, Folstein M. The telephone interview for cognitive status. Neuropsychiatry Neuropsychol Behav Neurol 1688;1:111–17. [Google Scholar]
- 12.Gobeille M, Bradley C, Goldstein JE, Massof R. Calibration of the Activity Inventory Item Bank: A patient-reported outcome measurement instrument for low vision rehabilitation. Transl Vis Sci Technol 2021;10:12. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Kennedy RS, Lane NE, Berbaum KS, Lilienthal MG. Simulator Sickness Questionnaire: An enhanced method for quantifying simulator sickness. Int J Aviat Psychol 1993;3:203–20. [Google Scholar]
- 14.Scheiman M, Mitchell LG, Cotter S, et al. A randomized clinical trial of treatments for convergence insufficiency in children. Arch Ophthalmol 2005;123:14–24. [DOI] [PubMed] [Google Scholar]
- 15.Camargo FR, Henson B. Conceptualising computerized adaptive testing for measurement of latent variables associated with physical objects. J Phys Conf Ser 2015;588:012012. [Google Scholar]
- 16.Jeter PE, Rozanski C, Massof R, et al. Development of the Ultra-Low Vision Visual Functioning Questionnaire (ULV-VFQ). Transl Vis Sci Technol 2017;6:11. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Bradley C, Massof RW. Method of successive dichotomizations: An improved method for estimating measures of latent variables from rating scale data. PLoS One 2018;13:e0206106. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Goldstein JE, Chun MW, Fletcher DC. Visual ability of patients seeking outpatient low vision services in the United States. JAMA Ophthalmol 2014;132:1169. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.How Luo G. 16,000 people used a smartphone magnifier app in their daily lives. Clin Exp Optom 2020;103:847–52. [DOI] [PubMed] [Google Scholar]
- 20.Costela FM, Reeves SM, Woods RL. An implementation of bubble magnification did not improve the video comprehension of individuals with central vision loss. Ophthalmic Physiol Opt 2021;41:842–52. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Lorenzini MC, Hämäläinen AM, Wittich W. Factors related to the use of a head-mounted display for individuals with low vision. Disabil Rehabil 2021;43:2472–86. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.