Skip to main content
. 2024 Apr 9;6(3):100938. doi: 10.1016/j.asmr.2024.100938

Table 2.

Failure Testing Results (n = 5 Biocomposite, n = 5 All-Suture)

Property Anchor Type
Pr > F
All-Suture
Biocomposite
Least Square Mean ± SEM
Load, N
 Yield 104.7 ± 6.5a 134.1 ± 6.5b 0.011
 Failure 196.8 ± 31.4 181.9 ± 31.4 0.647
Displacement, mm
 Yield 2.5 ± 1.1 3.0 ± 1.1 0.087
 Failure 8.8 ± 1.4 6.1 ± 1.4 0.255
Energy, J
 Yield 169.8 ± 85.7a 228.6 ± 85.7b 0.031
 Failure 1350 ± 390.5 716.8 ± 390.5 0.299
Stiffness, N/mm 42.1 ± 15.0 47.4 ± 15.0 0.366

NOTE. Values with different superscripts within an outcome measure are significantly different between anchor types.

Pr > F, p value of the F statistic; SEM, standard error of the mean.