Skip to main content
. 2024 Jul 8;14(13):1453. doi: 10.3390/diagnostics14131453

Table 2.

The values of variations between scan data (measured in micrometers).

Value Trios 5 CEREC Emerald S i700
the largest negative deviation −2000 µm −2000 µm −1999 µm −2000 µm
the largest positive deviation value 2000 µm 2000 µm 2000 µm 2000 µm
Median 11 µm 15 µm 14 µm 6 µm
Average Mean 22 µm 13 µm 29 µm 6 µm
Mean Absolute Deviation (MAD) 112 µm 127 µm 117 µm 114 µm
Root Mean Square Error (RMSE) 165 µm 179 µm 156 µm 174 µm
Standard Deviation (SD) 164 µm 179 µm 153 µm 174 µm
Variance 27 µm 32 µm 23 µm 30 µm
the average of the deviation values’ positive values 126 µm 129 µm 136 µm 117 µm
average of the deviation values’ negative values −97 µm −124 µm −95 µm −112 µm
(90–10)/2 (Trueness) 168 µm 192 µm 181 µm 161 µm
10% value of deviation values −141 µm −189 µm −146 µm −154 µm
90% value of deviation values 195 µm 195 µm 216 µm 168 µm
Tolerance 31.00% 26.79% 26.61% 29.11%