TABLE 5.

Number of HN autocontours (RPA) and manual contours (GT) that met clinical dose criteria.

Structure Metric, criteria RPA GT Number of data points
Brain Max dose < 70 Gy 54 54 54
BrainStem Max dose < 54 Gy 54 54 54
Chiasm Max dose < 54 Gy 5 4 5
Left cochlea Max dose < 35 Gy 50 51 54
Right cochlea Max dose < 35 Gy 50 50 54
Left eye Max dose < 54 Gy 37 38 38
Right eye Max dose < 54 Gy 37 37 38
Left kens Max dose < 7 Gy 50 50 53
Right kens Max dose < 7 Gy 50 50 53
Mandible Max dose < 70 Gy 45 45 54
Left OpticNrv Max dose < 54 Gy 10 9 10
Right OpticNrv Max dose < 54 Gy 10 9 10
Left parotid Mean dose < 26 Gy 36 36 54
Right parotid Mean dose < 26 Gy 38 38 54
SpinalCord Max dose < 45 Gy 54 54 54
zBrainStem_05 a Max dose < 54 Gy 50 50 52
zSpinalCord_05 a Max dose < 50 Gy 54 54 54

Abbreviations: GT, Ground Truth; HN, head and neck; RPA, Radiation Planning Assistant.

a

Structure with a 5‐mm margin.